Function Test
12
42
Z
Wave
runs a self-test that checks all internal components after it is
switched on.
An error message is shown in case of faults.
In addition, a function test can be performed as follows.
This test shall be made monthly or in case of doubt about the proper
function of the device.
Note:
Before performing the function test, check whether the handpiece and
the footswitch are connected correctly to the device.
Check the mains cable is connected properly to the device and the
network connection is established.
Function test
Testing
Switch on the device.
Depress the footswitch briefly – the fan and pulse generator will start
immediately, whereby the pulse generator has to operate at the
frequency indicated on the display (5 Hz as default value).
Note:
On conclusion of the test, switch off the device at the main switch.
If a treatment is to be performed immediately afterwards, set the required
treatment parameters and proceed as described in Chapter 4.