Manufacturer’s declaration of
Electromagnetic Compatibility
15
47
The main features of the Z
Wave
are as follows: interference-free delivery of pulses, interference-free control of all functions.
Uninterrupted operation is not required with the use intended.
Guidelines and manufacturer's declaration – electromagnetic interference immunity
The device Z
Wave
is intended for operation in the electromagnetic environment specified below. The customer or user of the
Z
Wave
should ensure that it is used in such an environment.
Interference
immunity tests
IEC 60601-test level
Compliance level
Electromagnetic environment - guidelines
Conducted RF
disturbance variables
according to IEC
61000-4-6
Radiated RF
disturbance variables
according to IEC
61000-4-3
3 V
effektive value
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
effektive value
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Portable and mobile radio equipment should
not be used any closer to the Z
Wave
, including
cables, than the recommended separation
distance calculated from the equation
applicable to the transmission frequency
.
Recommended separation distance:
d= 1.2
√
P
d= 0.35
√
P for 80 MHz to 800 MHz
d= 0,7
√
P for 800 MHz to 2.5 GHz
Where P is the rated power of the transmitter
in Watts (W) according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
According to an investigation in situ
a
, the field
strength of stationary radio transmitters should
be less than the compliance level at all
frequencies.
Interference may occur in the vicinity of
equipment which is marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz the higher frequency range is applicable.
NOTE 2 These guidelines may not be applicable in all cases. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.