Description
3.2 Operating principle
3
Page 16
Operating principle
3.2 Operating principle
The reModel high-frequency therapy device is equipped with three different
technologies: with a high-frequency function, with a low-level laser, and with the
function for vacuum massage. The handpiece for high-frequency and laser treatment
comprises three different sizes for different parts of the body and is primarily used to
relieve musculoskeletal pain.
The energy emission is controlled by pressing the button on the handpiece.
When the vacuum is switched on, air is aspirated through the solenoid valves
through underpressure in the handpiece.
o
High-frequency section: the HF circuit board generates a pulse with 2 MHz
and transmits this to the amplifier. The pulse is emitted via the high-voltage
transformer.
o
Laser function: This function is operated with charging capacity. The sealed
laser tubes generate the laser beam with a wavelength of 660 nm by
charging the high voltage into the sealed laser tubes, then the laser beam is
applied to the painful area.
o
Vacuum function: The air is aspirated to the lower section of the handpiece.
The air aspirated into the vacuum motor flows through the air filter and the
solenoid valve.