Manufacturer´s EMC declaration
18
Page 32
Guidelines and manufacturer declaration - electromagnetic interference resistance
The emField Pro device is intended for operation in the electromagnetic environment indicated
below. The emField Pro customer or user should ensure that it is used in such an environment.
Interference resistance
tests
IEC 60601- Test
level
Compliance level
Electromagnetic
environment - guidelines
Static electricity
discharge (SED) in
accordance with IEC
61000-4-2
± 6 kV contact
discharge
± 8 kV air
discharge
± 6 kV contact
discharge
± 8 kV air
discharge
Floors should be made of
wood or concrete or
covered in ceramic tiles. If
the floor is made of
synthetic material, the
relative humidity must be at
least 30%.
Rapid, transient
electrical disturbance
variables/bursts in
accordance with IEC
61000-4-4
± 2 kV for mains
cables
± 1 kV for input
and output
cables
± 2 kV for mains
cables
± 1 kV for input
and output cables
The quality of the supply
voltage should correspond
with a typical business or
hospital environment.
Surges
in accordance with IEC
6100-4-5
± 1 kV
differential mode
voltage
± 2 kV common
mode voltage
± 1 kV differential
mode voltage
± 2 kV common
mode voltage
The quality of the supply
voltage should correspond
with a typical business or
hospital environment.
Drops in voltage,
temporary interruptions
and fluctuations in the
supply voltage in
accordance with IEC
61000-4-11
<5% U
T
(>95% drop in
the U
T
for ½
period)
40% U
T
(60% drop in the
U
T
for 5 periods)
70% U
T
(30% drop in the
U
T
for 25
periods)
<5% U
T
(>95% drop in
the U
T
for 5
seconds)
<5% U
T
(>95% drop in the
U
T
for ½ period)
40% U
T
(60% drop in the
U
T
for 5 periods)
70% U
T
(30% drop in the
U
T
for 25 periods)
<5% U
T
(>95% drop in the
U
T
for 5 seconds)
The quality of the supply
should correspond with a
typical business or hospital
environment. If the emField
Pro user requires
continuous functionality
even if there are
interruptions in the power
supply, powering the
emField Pro using an
uninterrupted power supply
or battery is recommended.
Magnetic field at the
supply frequency (50/60
Hz) in accordance with
IEC 61000-4-8
3 A/m
3 A/m
Magnetic fields at the mains
frequency should
correspond with the typical
values found in business or
hospital environments.
Remark: U
T
is the mains AC voltage prior to applying the test level.