Device Set Up
6
Page 7
Caution
!
Caution
!
Switching the device
off
After using the emField
Pro
for a long time, the transducer arm may be stuck down or
up. In this case, the position of the transducer arm can be changed by adjusting the
gas spring pressure. To lower the arm supporter, turn the hexagonal wrench bolt in
the (
) direction as shown in the figure below, and turn it in the (
) direction to lift it
upwards.
To adjust the transducer arm use only the supplied hex key (24).
The device is switched off using the power switch (12).
In order to completely disconnect the device (all-phase) from the mains, remove the
mains cable.