Operation Instructions
8
Page 16
The parameter settings are:
- Wave
: Select one of the following waves:
Sinus; Triangle; Downward Triangle; Upward Triangle or
Flat by selecting the according symbol.
- L (Hz)
: Low limit of waveform (6-149 Hz). This frequency has to
be lower than H (Hz).
- H (Hz)
: High limit of waveform (7-150 Hz). This frequency has to
be higher than L (Hz).
- T1
: Duration of the stimulation phase (1 – 10 seconds).
The setting ranges are different according to the value of
the frequencies.
- OFF (Hz)
: Frequency during the Off phase (0 – 4 Hz).
- T2
: Duration of the Off phase. (4 – 10 seconds).
The setting ranges are different according to the value of
the frequencies.
- STR
: 1 – 100% (multiplying factor for the main magnetic field
intensity)
- Repetition
: Number of repetitions, 1 ~ 60 cycles.