Functional Test
15
Page 26
Self-troubleshooting
Symptom
What to do
References in the User Guide
Equipment does not
turn on.
Check if the power connector of the
equipment is properly connected.
• Chapter 6; device set up
Check if the power switch of equipment is
turned on.
• Figure 2; Rear of the device
Magnetic field is not
generated from
equipment.
Confirm if the LCD displays an ERROR
message.
• 17. Error Messages and
Troubleshooting
Check if the output strength is set by
turning the Turning Knob after pressing the
Start button
.
• 8. Operation Instructions
Over Temperature
message is displayed.
If transducer is in an upright position, it is
easy for the transducer to overheat.
Maintain a horizontal position as much as
possible.
• 17. Error Messages and
Troubleshooting
If the room temperature is too high,
disorder can be caused in the cooling.
Maintain the room temperature at less
than 25
℃
as much as possible.
• 17. Error messages and
Troubleshooting
In the following cases, stop the operation by cutting off the power to the equipment, and contact the
service center.
.
The main power switch spontaneously turns off.
The LCD screen of operation panel does not illuminate when power is turned off and then turned
on again.
When stimulation is not generated by transducer, even after intensity is increased.
The temperature icon on the screen blinks and the equipment is not operated.
Regular inspection for performance maintenance
In order to ensure safe use, be sure to have a regular, and at least early inspection by an authorized person
.