10
Adjustments, Settings and Menus
4.1
Detailed Menu Options
4.
Info
Displays system technical data.
No adjustments can be made in this option.
Automatic Service
Program S01
Only for use upon request of a technician.
Automatic Service
Program S02
Only for use upon request of a technician.
Device Configuration
Use this submenu to change the language used in the displays or to make technical
adjustments.
Use the arrow and Select Key to select the desired option.
Evaporator temperature displayed:
Select CHANGE (13, Figure 4) to change the state. If ON is selected, the Evaporator
temperature will be displayed in therapy Screen (18, Figure 4).
The displayed
temperature is NOT the AIR temperature.
OK
Start/ Stop
Evaporator temperature displayed
STORE
BACK
CHANGE
State: ON
Figure 10
Setting of warning threshold:
Use this Submenu to set the Temperature warning threshold, to display the warning
text in the Therapy screen and to activate the audio warning when the Evaporator
temperature reaches the setting warning threshold.
Audio warning:
Select CHANGE (13, Figure 4) to change the state. If ON is
selected, a beep sound will be generated for 2 Min. in 1 sec. Rhythm. Press Store
Key (10, Figure4) to apply and save the new setting.
OK
Start/ Stop
Audio warning
STORE
BACK
CHANGE
State: ON
Figure 11