15
Maintenance
5.
Service Elements: Back Panel
(8) Mains switch:
The mains power switch is located at the rear of the unit. Cryo
Mini
performs an
automatic self-test after mains switch is turned on (see Figure 2).
(7) Air filter:
The unit features an air filter (7, Figure 2) to remove coarse dust particles from the air
used for cooling and treatment. In normal use, it is sufficient to clean the filter, which is
located at the rear of the unit, with a vacuum cleaner. This should be carried out at
least every 150 hours of use, when the software issues a corresponding signal. If the
unit is operated in rooms with carpets or in other environments with a high level of
dust, it is recommended that the filter be cleaned more frequently.
Note: After cleaning the filter, the signal issued by the software is reset by pressing
the appropriate button mentioned in the display.
Power safety shut
down:
Cryo
Mini
is fitted with a 2-pole overload circuit breaker to protect the unit from power
supply irregularities. The circuit breaker is integrated into the Mains switch (8). If the
circuit breaker is triggered, the unit must be switched off before it can be reset.
Service Elements: Side Panel
(6) Defrosted water
container:
If Cryo
Mini
is switched off after use, or if the defrosting program is activated, the
cooling system warms up and creates condensation. The unit contains a Defrosted
Water Container , which holds about 1 liter and may be removed when full to empty it.
The container should be cleaned after emptying and then placed back in the device.
Cryo
Mini
should not be operated with the container removed.
!
!
!