21 |
Sesamoid
®
Plasty
User Guide
Troubleshooting of the Sesamoid Plasty
Camera Arm not Holding its Position
In some occasions, the camera may not hold its
position while orienting it to the operating field. The
following steps might fix this issue.
Note:
In order to calibrate the spring load of the
Camera Arm, the camera must be attached to it.
The Sesamoid Plasty must therefore be assembled
to proceed.
1. Using a 5.5 mm Allen key, tighten the set screw
located underneath the camera arm. Turning the
screw clockwise will increase the stiffness of the
arm.
2. To decrease the stiffness, loosen up the screw by
turning it counter clockwise.
3. Verify and adjust the stiffness of the camera arm
until it holds its position.
Location of the set screw that adjusts the stiffness of the camera arm
.
Communication Errors
When the Sesamoid Plasty Camera communication
error message is displayed
1. Verify that both LEDs are on and green.
3. Verify that the cable is plugged firmly into the
camera.
4. Verify that pins inside all connectors are not bent.
5. Make sure that the LEMO Connector sleeve is
tightly in place. Do not twist the cable at anytime.
6. Try changing USB ports.
7. Contact Zimmer CAS Customer Support if the error
is not resolved.
Bump Detection
When the internal bump detection message is
displayed, contact Zimmer CAS Customer Support.
Figure 12