16 |
Sesamoid
®
Plasty
User Guide
Technical Specifications
(cont.)
The Sesamoid Plasty is designed for use in the
operating room, with the following exceptions:
• Near portable and mobile RF communication
equipment that radiate electromagnetic fields as
those may cause performance degradation.
• In the presence of a flammable anesthetic mixture
with air, oxygen or nitrous oxide.
To ensure adequate communication between the
Sesamoid Plasty Camera and Computer, display of
the correct navigation values and to avoid latency in
the system, it is important to use the Sesamoid Plasty
in the recommended use environment.
Use of the Sesamoid Plasty outside of a suitable
environment can result in:
• Misleading/erroneous navigation information.
• Asynchronous information relative to current state
of instruments.
• Unintentional manipulations of the system.
CAUTION:
The emission characteristics of the
Sesamoid Plasty make it suitable for use in
industrial areas and hospitals (CISPR 11 class
A). The Sesamoid Plasty is not adequate for
use in a residential environment as it might not
offer adequate protection to RF communication
services.
Product Regulatory Notices
Wireless Communications
The Sesamoid Plasty Computer contains an RF
transmitter with the following characteristics:
• Transmission frequency: 2.4/5 GHz
• Network Standard: Compliant with IEEE 802.11n,
802.11g, 802.11b,802.11d, 802.11e, 802.11j and
802.11i
• Modulation: OFDM with BPSK, QPSK, 16 QAM, 64
QAM;DQPSK, CCK, G-FSK, π/4-DQPSK, 8-DPSK
• Supported Data Rates: IEEE 802.11a 6 - 54 Mbps,
IEEE 802.11b 1 - 11 Mbps, IEEE 802.11g 6 - 54
Mbps, IEEE 802.11n 6.5 - 300 Mbps
• Effective isotropic radiated power: 21 dBm