Allow the cut graft to accumulate in the pocket of the handpiece. Lift the handpiece away from 
the donor site to end the graft. Return the throttle to the SAFE position and carefully remove the 
graft. (See Fig. 10) 

• Method II  

Use tissue forceps to gently lift the graft as it emerges from the pocket area. Do not stretch or pull the 
graft as this causes irregular edges and nonuniform cuts. Lift the handpiece away from the donor site 
to end the graft. Return the throttle to the SAFE position. (See Fig. 11.)  

• Do not run the Zimmer Electric Dermatome without cutting for an extended time. Release the on/off 
lever and return the safety lock to the SAFE position between cuts to remove the graft. 

AFTER THE PROCEDURE

  

• Unplug from wall receptacle. 

 

• To remove the Zimmer Electric Dermatome connector plug from the power supply, pull outward on the 

outer sleeve of the plug to separate the plug from the receptacle.  

• Remove the used blade and dispose of properly in designated sharps container or per hospital protocol.  

• Remove excess body fluids and tissue with a disposable, non-shedding wipe and cover with a damp 
cloth. Body fluids and tissue should not be allowed to dry on instruments prior to cleaning. 

 

• Return and secure instrument and width plates to instrument case for transport and cleaning.  

BLADE REMOVAL 

 

• Loosen the width plate screws. Do not remove the screws.  

• To remove the width plate, hold both sides and lift. Do not lift the width plate from the front as this will 
cause contact with the blade and possible injury. 

 

• Remove the blade carefully lifting from the side.  

• Dispose of used blade in a sharps container or per hospital protocol. 

CLEANING PRECAUTIONS  

All subsequent cleaning and sterilization steps are facilitated by not allowing blood, tissue debris, or 
disinfectants to dry on used instruments.  

Handle the Zimmer Electric Dermatome carefully. Should it be inadvertently dropped or damaged, it 
should be serviced. 

 Refer to NEED TO ADHERE TO CARE REGIMEN section.  

Do not lubricate the Zimmer Electric Dermatome. Lubrication may cause extensive damage to the motor.  

Never immerse the dermatome in any solution. Some solutions will corrode the metal and delicate moving 
parts and also break down the internal lubricants.  

Never immerse the dermatome in liquid chemical disinfectant.  

Содержание Electric Dermatome

Страница 1: ...POWER SUPPLY I Physical Weight 2 6 lb 1 18 kg Length 9 12 in 23 2 cm Height 5 44 in 13 8 cm Width 6 69 in 17 0 cm II Electrical Power Requirements 100 240 V Auto Switching 150 VA 50 60 Hz Single Phase Power Output 14 5 V 4 3 A Maximum Protection Class Class 1 Degree of Protection Against Electrical Shock Type BF II HANDPIECE I Physical Weight 2 1 lb 95 kg Width 8 5 in 21 6 cm Length 5 2 in 13 2 cm...

Страница 2: ...FE position the safety lock on the throttle should be toward the blade end of the dermatome and only the word SAFE should be visible Handle the Zimmer Electric Dermatome carefully Should it be inadvertently dropped or damaged it should be serviced Do not use Use only Zimmer Dermatome Blades REF 00 8800 000 10 The Zimmer Dermatome Blade has been specifically designed and engineered for use with the...

Страница 3: ...he connector plug to the receptacle by rotating the plug while gently pushing inward When aligned push in until the plug clicks into the receptacle Note Older Zimmer Dermatomes and Power Supplies were equipped with a metal connector plug and receptacle The metal connectors have been upgraded to plastic connectors for ease of use The older metal connectors are not compatible with the plastic connec...

Страница 4: ...wever lubricating the donor site with sterile mineral oil may ease travel of the Zimmer Electric Dermatome Set control lever adjustment knob pointer to desired graft thickness Factory calibrations indicate 0 002 in 0 050 mm Do not insert any instrument between the blade and the control lever as this may damage or knick the blade causing a poor cut Further it may compromise the calibration of the i...

Страница 5: ...sue with a disposable non shedding wipe and cover with a damp cloth Body fluids and tissue should not be allowed to dry on instruments prior to cleaning Return and secure instrument and width plates to instrument case for transport and cleaning BLADE REMOVAL Loosen the width plate screws Do not remove the screws To remove the width plate hold both sides and lift Do not lift the width plate from th...

Страница 6: ...cautions for handling contaminated biohazardous materials should be observed Instruments should be cleaned within 30 minutes of use to minimize the potential for drying prior to cleaning Prepare cleaning agents at the use dilution and temperature recommended by the manufacturer Thoroughly scrub the instrument with a soft bristled brush and a pH neutral detergent If desired a neutral pH enzyme solu...

Страница 7: ...outgassing times are difficult to determine for lubricated powered instruments Never sterilize the power supply Disconnect the dermatome from the power supply before sterilization Place cleaned instruments in an instrument tray or fully perforated autoclave case The Zimmer Dermatome Autoclave Case REF 00 8801 003 00 is recommended See Fig 18 Do not kink or crimp the dermatomeʼs cord when closing t...

Страница 8: ... table 4 Disinfection steam sterilization parameters recommended by the World Health Organization WHO for reprocessing instruments where there is concern regarding TSE CJD contamination 5 For Universal Instrument Cases without defined load configurations 6 Sea level 7 AAMI AORN steam sterilization cycles with longer times than those listed are also acceptable 8 Medical grade steam sterilization co...

Страница 9: ...00 8802 003 00 00 8802 004 00 00 8803 000 00 REF 00 8821 001 00 Description Power Supply includes Fuse drawer 5 mm x 20 mm Fuses 1 25 A QTY 2 Fuses 800 mA QTY 2 Power Cord REF 00 8821 006 00 Description Autoclave Case REF 00 8801 003 00 Description 1 in 2 5 cm Width Plate REF 00 8802 001 00 Description 1 5 in 3 8cm Width Plate REF 00 8802 015 00 Description 2 in 5 1 cm Width Plate REF 00 8802 002 ...

Страница 10: ...tions Have unit serviced Dermatome not operating Power light on Check blade fit Check handpiece connection to power supply Have unit serviced Dermatome operating too slowly Check dermatome for damage Check blade for damage Have unit serviced Erratic speed changes Check handpiece connection to power supply Have unit serviced SERVICE INFORMATION Annual calibration checks are strongly recommended to ...

Страница 11: ...A Autoclave Case B Power Supply C Screwdriver D Width Plates E Handpiece ...

Страница 12: ...A Front B Mains Switch C Connector Receptacle D Power Entry Module E Back A On Off Safety Lock B On Off Lever C Handpiece D Connector Plug E Thickness Control Lever ...

Страница 13: ...A Screwdriver B Width Plate C Blade D Drive Pin Hole E Drive Pin ...

Страница 14: ......

Страница 15: ......

Страница 16: ...METHOD I A Handpiece B Pocket METHOD II A Handpiece B Forceps ...

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ......

Страница 21: ......

Страница 22: ......

Отзывы: