POWER SOURCE  

The unit is shipped with a 0.25 in. (6.35 mm) x 1.25 in. (31.75 mm) fuse drawer installed. A separate fuse 
drawer is supplied for 0.197 in. (5 mm) x 0.787 in. (20 mm) fuses. To replace input fuses, disconnect the 
power supply from the power source. Remove the fuse drawer by lifting up on the tab using a tool and 
pulling the drawer outward. (See Fig. 6.) Replace the fuse drawer with the correct fuses installed. 

FUSES  

Input fuses are located in the fuse drawer in the power entry module at the rear of the power supply. 250 
V ~ 1.25 A time lag/delay. 

WARNINGS AND PRECAUTIONS 

To avoid serious injury to the patient and operating staff while the Zimmer Electric Dermatome is in use, 
the user must be thoroughly familiar with its function, application, and instructions for use.  

To avoid injury, use extreme caution when handling the blade or when handling the dermatome with the 
blade installed. 

 Use caution when inserting blade to avoid nicking it, which may result in an uneven cut. To avoid blade 
damage, place dermatome blade side up when not in use. 

 The throttle must be in the SAFE position before changing blades, when connecting power to the 
instrument, or when the instrument is not in use. Accidental activation of the instrument during these 
procedures may injure the patient or operating staff. To ensure that the instrument is in the SAFE 
position, the safety lock on the throttle should be toward the blade end of the dermatome and only the 
word SAFE should be visible.  

Handle the Zimmer Electric Dermatome carefully. Should it be inadvertently dropped or damaged, it 
should be serviced. Do not use. 

 Use only Zimmer Dermatome Blades (REF 00-8800-000-10). The Zimmer Dermatome Blade has been 
specifically designed and engineered for use with the Zimmer Electric Dermatome. Other blades may not 
fit properly in the dermatome and may cause serious injury. Use of non-Zimmer Dermatome Blades can 
cause the dermatome to take grafts deeper than what the user has selected. Never connect the Zimmer 
Electric Dermatome Handpiece to any source other than the Zimmer Electric Dermatome Power Supply. 
It has been factory calibrated to provide optimum cutting performance and maximum safety. It is also 
designed to meet or exceed specific medical electrical safety standards. 

Ground reliability can only be achieved when the power cord is connected to a receptacle marked 
“HOSPITAL GRADE”. Use only the power cord marked “HOSPITAL GRADE” supplied or one that 
complies with all local and electrical requirements but does not exceed 10 feet (3.05 meters).  

For continued protection against fire hazard, replace only with the same type and rating of fuse. Refer to 
FUSES section.  

Possible explosion hazard exists if this instrument is used in the presence of flammable anesthetics or 
gasses.  

Never sterilize the power supply. Disconnect the dermatome handpiece from the power supply before 
sterilization.  

Содержание Electric Dermatome

Страница 1: ...POWER SUPPLY I Physical Weight 2 6 lb 1 18 kg Length 9 12 in 23 2 cm Height 5 44 in 13 8 cm Width 6 69 in 17 0 cm II Electrical Power Requirements 100 240 V Auto Switching 150 VA 50 60 Hz Single Phase Power Output 14 5 V 4 3 A Maximum Protection Class Class 1 Degree of Protection Against Electrical Shock Type BF II HANDPIECE I Physical Weight 2 1 lb 95 kg Width 8 5 in 21 6 cm Length 5 2 in 13 2 cm...

Страница 2: ...FE position the safety lock on the throttle should be toward the blade end of the dermatome and only the word SAFE should be visible Handle the Zimmer Electric Dermatome carefully Should it be inadvertently dropped or damaged it should be serviced Do not use Use only Zimmer Dermatome Blades REF 00 8800 000 10 The Zimmer Dermatome Blade has been specifically designed and engineered for use with the...

Страница 3: ...he connector plug to the receptacle by rotating the plug while gently pushing inward When aligned push in until the plug clicks into the receptacle Note Older Zimmer Dermatomes and Power Supplies were equipped with a metal connector plug and receptacle The metal connectors have been upgraded to plastic connectors for ease of use The older metal connectors are not compatible with the plastic connec...

Страница 4: ...wever lubricating the donor site with sterile mineral oil may ease travel of the Zimmer Electric Dermatome Set control lever adjustment knob pointer to desired graft thickness Factory calibrations indicate 0 002 in 0 050 mm Do not insert any instrument between the blade and the control lever as this may damage or knick the blade causing a poor cut Further it may compromise the calibration of the i...

Страница 5: ...sue with a disposable non shedding wipe and cover with a damp cloth Body fluids and tissue should not be allowed to dry on instruments prior to cleaning Return and secure instrument and width plates to instrument case for transport and cleaning BLADE REMOVAL Loosen the width plate screws Do not remove the screws To remove the width plate hold both sides and lift Do not lift the width plate from th...

Страница 6: ...cautions for handling contaminated biohazardous materials should be observed Instruments should be cleaned within 30 minutes of use to minimize the potential for drying prior to cleaning Prepare cleaning agents at the use dilution and temperature recommended by the manufacturer Thoroughly scrub the instrument with a soft bristled brush and a pH neutral detergent If desired a neutral pH enzyme solu...

Страница 7: ...outgassing times are difficult to determine for lubricated powered instruments Never sterilize the power supply Disconnect the dermatome from the power supply before sterilization Place cleaned instruments in an instrument tray or fully perforated autoclave case The Zimmer Dermatome Autoclave Case REF 00 8801 003 00 is recommended See Fig 18 Do not kink or crimp the dermatomeʼs cord when closing t...

Страница 8: ... table 4 Disinfection steam sterilization parameters recommended by the World Health Organization WHO for reprocessing instruments where there is concern regarding TSE CJD contamination 5 For Universal Instrument Cases without defined load configurations 6 Sea level 7 AAMI AORN steam sterilization cycles with longer times than those listed are also acceptable 8 Medical grade steam sterilization co...

Страница 9: ...00 8802 003 00 00 8802 004 00 00 8803 000 00 REF 00 8821 001 00 Description Power Supply includes Fuse drawer 5 mm x 20 mm Fuses 1 25 A QTY 2 Fuses 800 mA QTY 2 Power Cord REF 00 8821 006 00 Description Autoclave Case REF 00 8801 003 00 Description 1 in 2 5 cm Width Plate REF 00 8802 001 00 Description 1 5 in 3 8cm Width Plate REF 00 8802 015 00 Description 2 in 5 1 cm Width Plate REF 00 8802 002 ...

Страница 10: ...tions Have unit serviced Dermatome not operating Power light on Check blade fit Check handpiece connection to power supply Have unit serviced Dermatome operating too slowly Check dermatome for damage Check blade for damage Have unit serviced Erratic speed changes Check handpiece connection to power supply Have unit serviced SERVICE INFORMATION Annual calibration checks are strongly recommended to ...

Страница 11: ...A Autoclave Case B Power Supply C Screwdriver D Width Plates E Handpiece ...

Страница 12: ...A Front B Mains Switch C Connector Receptacle D Power Entry Module E Back A On Off Safety Lock B On Off Lever C Handpiece D Connector Plug E Thickness Control Lever ...

Страница 13: ...A Screwdriver B Width Plate C Blade D Drive Pin Hole E Drive Pin ...

Страница 14: ......

Страница 15: ......

Страница 16: ...METHOD I A Handpiece B Pocket METHOD II A Handpiece B Forceps ...

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ......

Страница 21: ......

Страница 22: ......

Отзывы: