(Date / Name): 16.04.2009 Sc
Drawing no.: 12280-0701-00
Page 23 of 24
Name: TR 800 WEB
ZIEHL industrie-elektronik GmbH + Co KG, Daimlerstr.13, D-74523 Schwäbisch Hall, Tel.: +49 791 504-0, Fax: -56, email: [email protected]
Reset input Y1/Y2
app. DC 18 V / 3,5 mA
Sensor connection:
Measuring-cycle / time
< 3 s
Pt 100, Pt 1000 according to EN 60751:
Measuring
range °C
Short-circuit
Ohm
Break
Ohm
Sensor resistance
+ line resistance Ohm
Sensor
min
max
<
>
max
Pt 100
-199
860
15
400
500
Pt 1000 -199
860
150
4000
4100
KTY 83
-55
175
150
4000
4100
KTY 84
-40
150
150
4000
4100
Tolerance
±0,5 % of measured value ±0,5 K (KTY ±5 K)
Sensor current
≤
0,6 mA
Temperature drift
<0,04°C/K
Thermocouples according to EN 60 584, DIN 43 710:
Measuring range °C
Precision
Typ
min
Max
B
0
1820
±2 °C
T > 300°C
E
-270
1000
±1 °C
J
-210
1200
±1 °C
K
-200
1372
±2 °C
L
-200
900
±1 °C
N
-270
1300
±2 °C
R
-50
1770
±2 °C
S
-50
1770
±2 °C
T
-270
400
±1 °C
Temperature drift
< 0,01 % / K
Measuring error of the sensor wire
+0,25 µV /
Ω
Reference junction
±5
°C
Voltage- / Current input
Input
resistance
Maximum
Input signal
Precision
(from
FullScale)
0 – 10 V
12 k
Ω
27 V
0,1 %
0/4-20 mA
18
Ω
100 mA
0,5 %
Temperature drift
< 0,02 %/K
Measuring of resistance:
Precision 0,0 … 500,0
Ω
0,2 % of measured value ± 0,5
Ω
Precision 0…30,00 k
Ω
0,5 % of measured value ± 2
Ω
Sensor-current
≤
0,6 mA