Display
Liquid crystal display section (65 mm x 30 mm) with analog indication and
digital display and with display of the unit of measured quantity, function
and various special functions.
Analog :
Indication
LCD scale with pointer
Scale length
55 mm on V
and A
, 47 mm on all other ranges
Graduation
5...0...
30 with 35 scale divisions on
,
0...30 with 30 scale divisions on all other ranges
Polarity indication
with automatic change-over
Overrange indication by triangle (13)
Sampling rate
20 readings/s, on W ; 10 readings/s
Digital:
Display/Height of numer. 7-segment numerals/15mm
Number of digits
3
3/4
digit
3100 counts
Overrange
"OL"is displayed.
Polarity indication
" - "sign is displayed, when the positive pole is at "
"
Sampling rate
2 reading/s, on W and
0
C:
1 reading/s
Power supply
Battery
9V at cell battery; manganese-dioxide cell according
to IEC 6 F 22, alkaline-manganese cell according to
IEC 6 LR 61 or suitable
NiCd storage battery.
Lifespan
With alkaline-manganese cell:
approx. 750 hours on V
, A
approx. 200 hours on V
~
,
A
~
(12S...15S)
approx. 150 hours on V
~
,
A
~
(16 S)
When operating with interface: times x 0.7
Battery test
automatic display of the "
" symbol, when the
battery voltage drops below approx. 7 V.
Electrical safety
Protection class
II according to IEC 348 / DIN VDE 0411 and
IEC 1010-1/EN61010-1/VDE0411-1
Overvoltage category II
III
Nominal voltage
1000V
600V
Contamination degree 2
2
Nominal insulation voltage 1000V according to IEC 348 / DIN VDE 0411
Test voltage
6kV~according to IEC 348/DIN VDE0411
EMC
Electromagnetic compatibility
Emission
EN50081-1:1992 / EN55022:1987 class B
Immunity
EN50082-1:1992 / IEC801-2:1991 8kV AD
/ IEC801-3:1984 3V/m
/ IEC801-4:1988 0.5kV
Fuses
Fuse for upto 300mA ranges
Fast blowing 1.6 A / 500V; 6.3mm X 32mm; rating 20kA with
500V
~
and ohmic load; in conjuction with power diodes, protects
all current measuring ranges upto 300mA.
Fuse for upto 10A ranges
16 A / 600V; 10mm x 38mm; rating 100 kA with 600V
~
and
ohmic load; protects the 3A and 10A ranges upto 600V; see
"18. maintenance" for manufacturers and types of fuses.
6
divisions so that variations of the measured values around "zero"can be
observed exactly. When the measured value exceeds the range of
indication, the left triangle (16) is shown before the polarity of the analog
indicator switches over after approximately 0.7s. The over range indication
on the measuring range (>3099 digits, on the range
>1999) is shown
by the right triangle (13).
6. ’’DATA’’hold facility
The DATA function allows to automatically hold the measured values.
This is particularly useful, for instance, when connecting the probes to
the measuring point requires full attention. When the measured value is
applied and the ’’condition’’ according to the table shown below is met,
the meter holds the measured value on the digital display and emits a
sound signal. The probes can now be removed from the measuring point
and the measured value on the digital display (9) can be read. When the
measured value falls below the limit specied in the table, the meter is
reactivated for a new storage.
The analog indication is not inuenced by the DATA
hold, The actual
measured value can still be noted / read. Note that with a held digital
display, the location of the decimal point is also held. With autoranging
selected, the measuring range of the analog indicator is no longer known.
1) Reactivated by falling below the specied limits of the measured value.
2) With the exception of the ranges 30 mV and 300 mV.
3) 240/280 digits when the ranges 3 mA for
RISH
Multi
12S/13S and 300 mA, 30
mA, 3 A for
RISH
Multi
14S...16S are selected according to section4.1.
As long as the DATA hold function is active, manual range selection is
not possible. The DATA hold function is switched OFF, when,
The "DATA” pushb utton (3)
is pressed for approx. 1s. This is
acknowledged by 2 sound signals.
The function selector switch (6) is operated or
The multimeter is turned OFF and ON again.
7.
Minimum value and Maximum value ’’MIN/MAX" storage facility.
With the MIN/MAX function, you can hold the minimum and the maximum
measured value which was applied to the input of the multimeter after
activating MIN/MAX function. The most important application is the
determination of the minimum and the maximum value for long-term
monitoring of measured quantities. MIN/MAX does not inuence the
analog indication The actual measured value can still be noted/read.
Apply the measured quantity to the meter and select the measuring
range prior to activating the MIN/MAX function.
With the function activated, you can select the measuring ranges only
manually, if you switch to another range, the stored MIN/MAX values are
cleared.
Function
DATA
Activate
Store
Reactive
1)
Reset
Short
Long
DATA
MIN/MAX
(3)
Condition
Measuring
Ranges
>280
>24
3)
< OL
>280
< 280
< 24
3)
OL
< 280
dis-
played
stored
mea-
sured
value
Cleared
ashes
dis-
played
ashes
Cleared
1 x
1 x
2 x
Meteracknowledgement
Display
DATA
Sound
Signal
Meas.
Value
digital
Limit of
Measured
Values
(digits)
V
2)
A
W
F,Hz,%
V
2)
A
W
F,Hz, %