23

Display

Liquid crystal display section (65 mm x 30 mm) with analog indication and
digital display and with display of the unit of measured quantity, function
and various special functions.

Analog :

Indication

LCD scale with pointer

Scale length

55 mm on V

and A

, 47 mm on all other ranges

Graduation

 5...0...

30 with 35 scale divisions on

,

0...30 with 30 scale divisions on all other ranges

Polarity indication

with automatic change-over

Overrange indication by triangle (13)
Sampling rate

20  readings/s, on  W ; 10 readings/s

Digital:

Display/Height of numer. 7-segment numerals/15mm
Number of digits

3

3/4

 digit 

 3100 counts

Overrange

"OL"is displayed.

Polarity indication

" - "sign is displayed, when the positive pole is at "

"

Sampling rate

2  reading/s, on  W and

 0

C: 

1 reading/s

Power supply

Battery

9V  at cell battery; manganese-dioxide cell according
to IEC 6 F 22, alkaline-manganese cell according to
IEC 6 LR 61 or suitable
NiCd  storage battery.

Lifespan

With alkaline-manganese cell:
approx. 750 hours on V

, A

approx. 200 hours on V 

~

,

 

~

 (12S...15S)

approx. 150 hours on V 

~

,

 

~

 (16 S)

When operating with interface: times x 0.7

Battery test

automatic display of the "

" symbol, when the

battery voltage drops below approx. 7 V.

Electrical safety

Protection class

II  according to IEC 348 / DIN VDE 0411 and

IEC 1010-1/EN61010-1/VDE0411-1

Overvoltage category II

III

Nominal voltage

1000V

600V

Contamination degree 2

2

Nominal insulation   voltage 1000V according to IEC 348 / DIN VDE 0411
Test voltage

6kV~according to IEC 348/DIN VDE0411

EMC

Electromagnetic compatibility

Emission

EN50081-1:1992 / EN55022:1987 class B

Immunity

EN50082-1:1992 / IEC801-2:1991 8kV AD

/ IEC801-3:1984 3V/m
/ IEC801-4:1988 0.5kV

Fuses
Fuse for upto 300mA ranges

Fast blowing 1.6 A / 500V; 6.3mm X 32mm; rating 20kA with
500V

~

 and ohmic load; in conjuction with power diodes, protects

all current measuring ranges upto 300mA.

Fuse for upto 10A ranges

16 A / 600V; 10mm x 38mm; rating 100 kA with 600V

 ~ 

and

ohmic load; protects the 3A and 10A ranges upto 600V; see
"18. maintenance" for manufacturers and types of fuses.

6

divisions so that variations of the measured values around "zero"can be
observed exactly. When the measured value exceeds the range of
indication, the left triangle (16) is shown before the polarity of the analog
indicator switches over after approximately 0.7s. The over range indication
on the measuring range (>3099 digits, on the range

>1999) is shown

by the right triangle (13).

6. ’’DATA’’hold facility

The DATA  function allows to automatically hold the measured values.
This is particularly useful, for instance, when connecting the probes to
the measuring point requires full attention. When the measured value is
applied and the ’’condition’’ according to the table shown below is met,
the meter holds the measured value on the digital display and emits a
sound signal. The probes can now be removed from the measuring point
and the measured value on the digital display (9) can be read. When the
measured value falls below the limit specied in the table, the meter is
reactivated for a new storage.
The analog indication is not inuenced by the DATA

hold, The actual

measured value can still be noted / read. Note that with a held digital
display, the location of the decimal  point is also held. With autoranging
selected, the measuring range of the analog indicator is no longer known.

1) Reactivated by falling below the specied limits of the measured value.
2) With the exception of the ranges 30 mV and 300 mV.
3) 240/280 digits when the ranges 3 mA for 

RISH 

Multi 

12S/13S and 300 mA, 30

mA, 3 A for 

RISH

 Multi

 14S...16S are selected according to section4.1.

As long as the DATA  hold function is active, manual range selection is
not possible. The DATA  hold function is switched OFF, when,

The  "DATA” pushb utton (3) 

is pressed for approx. 1s. This is

acknowledged by 2 sound signals.
The function selector switch (6) is operated or
The multimeter is turned OFF and ON again.

7.

 

Minimum value and Maximum value ’’MIN/MAX"  storage facility.

With the MIN/MAX function, you can hold the minimum and the maximum
measured value which was applied to the input of the multimeter after
activating MIN/MAX function. The most important application  is the
determination of the minimum and the maximum value for long-term
monitoring of measured quantities. MIN/MAX does not inuence the
analog indication The actual measured value can still be noted/read.
Apply the measured quantity to the meter and select the measuring
range prior to activating the MIN/MAX function.
With the function activated, you can select the measuring ranges only
manually, if you switch to another range, the stored MIN/MAX values are
cleared.

Function

DATA

Activate

Store

Reactive

1)

Reset

Short

Long

DATA

MIN/MAX

(3)

Condition

Measuring

Ranges

>280
>24

3)

< OL
>280

< 280
< 24

3)

OL

< 280

dis-

played

stored

mea-

sured
value

Cleared

ashes

dis-

played

ashes

Cleared

1 x

1 x

2 x

Meteracknowledgement

Display

DATA

Sound

Signal

Meas.

Value

digital

Limit of

Measured

Values

(digits)

V

 

2)

A

 

W

F,Hz,%

V

2)

A

 

W

F,Hz, %

Содержание MULTI 12S

Страница 1: ...Printed in India Subject to change without Notice Genius 04 05 2T User Manual ZIEGLER 12S 16S Analog Digital Multimeter Multi ...

Страница 2: ...10 A fused 1000 V max 7 10 0 20 30 ON OFF TRMS MAN MAX DATA MIN MAX AUTO MAN F mA 0 C A F W Hz V V V mA 0 C A 8 9 10 11 12 16 15 14 13 18 17 19 VA MIN MAX DC AC ON 8 8 8 8 10 0 20 30 MAN DATA KW MW kHz 0 C mF nF mm 5 19 Servicing When you need service please contact ...

Страница 3: ...measurement Rish multi 15S and 16S 14 14 Duty cycle measurement Rish multi 15S and 16S 15 15 Temperature measurement 15 16 Computer interface via RS 232C 16 17 Speci cations 18 18 Maintenance 26 18 1 Battery 26 18 2 Fuses 27 18 3 Case 27 19 Servicing 29 1 Liquid crystal display 2 ON OFF pushbutton 3 Pushbutton for data hold and MIN MAX storage functions 4 Pushbutton for manual range selection 5 Mu...

Страница 4: ...R Multi 12S 16S are manufactured and tested in compliance with the safety standard DIN VDE0411 and IEC 1010 1 DIN EN61010 1 VDE0411 1 In case of incorrect use or careless handling the safety of both user and multimeter is not assured For proper use and safe handling it is absolutely necessary to read and understand the operating instructions before using the meter For your safety and for protectio...

Страница 5: ...x 0 7 s max 9 s max 2 5 s from 0 to 80 of upper range limit from oo to 50 of upper range limit from 0 to 50 of upper range limit 3 Repair replacement of parts When opening the meter live parts may be exposed Therefore the meter must be disconnected from the measuring circuit prior to opening its case for repair or replacement of parts If repair cannot be avoided unless the meter is opened and live...

Страница 6: ...FF Press the ON OFF pushbutton 2 4 Function and range selection The function selector switch 6 is coupled with the Automatic terminal Blocking System ABS which allows access only to two correct sockets for each function Prior to switching to the mA or A functions or from the mA or A functions remove the test lead from the corresponding socket When the test leads are plugged in the terminal blockin...

Страница 7: ...section 4 1 4 3 Manual range section You can switch OFF autoranging and select the ranges manually according to the table on the following page Manual mode is switched OFF when pushbutton AUTO MAN is pressed 4 for approximately 1s when the function selector switch 6 is operated or when the meter is turned OFF and ON again When switching back to autoranging from 30 mV or 300 mV ranges 3 V range is ...

Страница 8: ...999 is shown by the right triangle 13 6 DATA hold facility The DATA function allows to automatically hold the measured values This is particularly useful for instance when connecting the probes to the measuring point requires full attention When the measured value is applied and the condition according to the table shown below is met the meter holds the measured value on the digital display and em...

Страница 9: ...djustment on the 30 mV measuring range Connect the test leads to the meter and join the free ends After having selected the measuring range brie y press the yellow multi function pushbutton 5 The meter acknowledges zero setting by a sound signal the LCD shows 00 00 1 digit and the decimal point ashes The displayed voltage at the instant the pushbutton is pressed is used as reference value max 200 ...

Страница 10: ...1 0 29 500 V DC AC eff rms sine 10 min 3 kHz 1200 V 30 kHz 300V 100kHz 30 V 500 V DC AC eff rms sine continu ously 10 min 7 Range 3 V U E 1 5 V eff rms 100 V eff rms 30 V UE 15 V eff rms 300 V eff rms 300V UE 150 V eff rms 1000 V eff rms 8 On the range 3V square wave signal positive on one side 5 15 V f const not 163 84 Hz or integral multiple 9 Without sensor 8 Voltagemeasurement Voltagemeasureme...

Страница 11: ...ed alternate switching takes place between DC and AC on the ZIEGLER Multi12S 15S or between DC and DC AC on the ZIEGLER Multi16S The change over is acknowledged by a sound signal The symbols DC and AC 11 are displayed as per selected function on the LCD When selecting a range with the function selector switch 6 the DC function is always set by default When pressing the yellow multi function pushbu...

Страница 12: ... 300 Digit Contin uously 0 75 3 10 Digit 1 0 5 10 Digit 1 0 5 10 D 1 0 2 0 25 2 1 0 5 10 Digit 1 0 2 1 0 5 10 Digit 1 0 2 1 5 2 10 Digit 1 5 2 10 Digit 1 5 2 10 Digit 1 5 2 10 Digit 1 5 2 10 Digit 0 36 A 5 0 36 A 5 6 6 6 6 0 36 A 5 0 5 3 4 0 5 3 0 4 1 0 4 1 0 4 1 0 6 1 2 0 1 0 25 1 0 5 5 10 D 0 5 2 0 5 5 10 D 0 5 2 1 0 5 10 D 1 0 2 1 5 4 10 D 1 5 4 10 D 1 75 4 10D 0 4 34 0 4 3 0 2 1 0 2 1 0 2 1 0 ...

Страница 13: ...tware MS WINDOWS 3 0 or higher Hardware IBM compatible PC with 1 MB main memory for the WINDOWS version A VGA or EGA monitor A hard disk with 2 MB free storage space A 3 5 disk drive for disks with 1 44 MB storage capacity A MICROSOFT compatible mouse If you wish to make hardcopy a printer supported by WINDOWS ...

Страница 14: ...ic load Caution The measuring ranges 3A and 16 A of ZIEGLER Multi13S are not protected by a fuse A blown fuse is signalled on the LCD the instant a measured quantity having a voltage of more than 4 V is applied to the corresponding connection sockets Then the digital display 9 shows the word FUSE After a fuse has blown eliminate the cause of the overload before using the meter again Replacement of...

Страница 15: ...unction selector switch 6 15 14 Duty cycle measurement ZIEGLER Multi 15S and 16S With duty cycle measurement we can determine the ratio of pulse duration to cycle time of recurring square wave signals Set the function selector switch 6 to V or V Connections are made in the same way as for voltage measurement See foot note 8 on page 21 Brie y press the yellow multi function pushbutton 5 twice The m...

Страница 16: ...Cll format for further processing For the use of ZIEGLER com 100 the following requirements must be met ON 13 Zero adjustment on the 30 Wmeasuring range When measuring small resistance values on the 30 W range you can eliminate the resistance of the leads and contact resistance by zero adjustment Connect the test leads to the multimeter and join the free ends Brie y press the yellow multi function...

Отзывы: