background image

24

Inuence

quantity

RangeofInuence

Measured

quantity/

Measuring range

Variation 

1)

+ (...% of rdg. + ... digits)

12 S...14S

15 S

16 S

Temperature

Frequency

of the

measured

quantity

0

C

+ 21 

0

C

and

+ 25 

 0

C... + 40 

0

C

15 Hz ... < 30 Hz

30 Hz... < 45 Hz

> 65 Hz... 400 Hz
> 400 Hz... 1 kHz
> 1 kHz... 20 kHz

15 Hz... < 30 Hz

30 Hz.. < 45 Hz

> 65 Hz... 1 kHz
15 Hz... < 30 Hz

30 Hz ... < 45 Hz

> 65 Hz... 1 kHz

Crest.

1... 3

factor CF

> 3...5

30/300 mV 

1.0 + 3

1.0 + 1

3... 300 V 

0.15 + 1

0.1 + 1

1000 V

 

0.2 + 1

0.1 + 1

~

0.4 + 2

0.3 + 2

300 

m

A

2) 

...

0.5 + 1

0.15 + 1

300 mA

 

3A / 10 (16) A

 

0.5 + 1

~

0.75 + 1

0.75+ 3

30 

W

2)

0.15 + 2

300 

W

0.25 + 2

0.15 + 2

3K  

W - 3MW

0.15 + 1

0.1 + 1

30 M 

W

1.0 + 1

0.6 + 1

30 nF

2)

 - 3 

m

F

-

0.5 + 2

30 

m

F

-

2.0 + 2

Hz

-

0.5 + 1

%

-

+ 5 Digit

-  200 ... + 200 

0

C

0.5 K + 2

+ 200 ... + 850 

0

C

0.5 + 2

3 ... 300 V

 ~

1000 V 

~

~

V

 ~

 

4)

 , A 

~

 

4)

-

-

1.0 + 3

-

-

0.5 + 3

2.0 + 3

0.5 + 3

2.0 + 3

1.0 + 3

-

-

2.0 + 3

-

-

1.0 + 3

-

-

0.5 + 3

3.0 + 3

2.0 + 3

-

-

1.0 + 3

-

-

0.5 + 3

2.0 + 3

3.0 + 3

-

-

+ 1 % of rdg.

-

-

+ 3 % of rdg.

Voltage measurement

Current measurement

Inuence Quantities and Variations

0

500 V

1000V

CF

5

4

3

2

1

0

0

1000

2000

3000

CF

5

4

3

2

1

0

The permissible crest factor CF of the AC quantity to be measured is a function of the
displayed value :

Wave form

of the

measured

quantity

5

The meter switches automatically to :
  the next higher range

at     +  (3099 1 digit)

  the next lower range

at + (240/280 digits - 1 digit)

   from the 300mA 

 to the 3mA 

 range

at            + (24 digits - 1 digit)

provided the 300 mA, 30 mA and 3 A ranges are not selected according
to section 4.1

4.3 Manual range section

You can switch OFF autoranging and select the ranges manually
according to the table on the following page.
Manual mode is switched OFF when pushbutton AUTO/MAN is pressed
(4) for approximately 1s, when the function selector switch (6) is operated,
or when the meter is turned OFF and ON again.
When switching back to autoranging from 30 mV

or 300 mV

 ranges,

3 V

range is automatically selected.

1) When these measuring ranges are selected
2) except for

 

ZIEGLER

 

Multi

12S/13S

3) except for 

ZIEGLER

 Multi

 14S...16S

5. Liquid crystal display

5.1Digital display

The digital display (9) shows the measured value with correct location of decimal
point and sign. The selected measuring Unit (12) and the function (11)
are simultaneously  displayed. When measuring DC quantities, a minus sign
appears in front of the digits, when the positive pole of the measured
quantity is applied to the "

"

 

input terminal. When upper range limit 3099

( on the range 

:1999), is exceeded then "OL" is displayed.

With V, A  and W measurements, the digital display is updated two times
per second.

5.2  Analog indication

The analog indication with pointer presentation gives the dynamic
response of a moving-coil movement and is updated 20 times per second,
when measuring V, A and  W. Analog indication is of particular advantage
when observing variations of measured values and for calibration
procedures.
The analog indicator has its own polarity indication. When measuring
DC quantities, the analog scale (15) has a negative range of 5 scale

Содержание MULTI 12S

Страница 1: ...Printed in India Subject to change without Notice Genius 04 05 2T User Manual ZIEGLER 12S 16S Analog Digital Multimeter Multi ...

Страница 2: ...10 A fused 1000 V max 7 10 0 20 30 ON OFF TRMS MAN MAX DATA MIN MAX AUTO MAN F mA 0 C A F W Hz V V V mA 0 C A 8 9 10 11 12 16 15 14 13 18 17 19 VA MIN MAX DC AC ON 8 8 8 8 10 0 20 30 MAN DATA KW MW kHz 0 C mF nF mm 5 19 Servicing When you need service please contact ...

Страница 3: ...measurement Rish multi 15S and 16S 14 14 Duty cycle measurement Rish multi 15S and 16S 15 15 Temperature measurement 15 16 Computer interface via RS 232C 16 17 Speci cations 18 18 Maintenance 26 18 1 Battery 26 18 2 Fuses 27 18 3 Case 27 19 Servicing 29 1 Liquid crystal display 2 ON OFF pushbutton 3 Pushbutton for data hold and MIN MAX storage functions 4 Pushbutton for manual range selection 5 Mu...

Страница 4: ...R Multi 12S 16S are manufactured and tested in compliance with the safety standard DIN VDE0411 and IEC 1010 1 DIN EN61010 1 VDE0411 1 In case of incorrect use or careless handling the safety of both user and multimeter is not assured For proper use and safe handling it is absolutely necessary to read and understand the operating instructions before using the meter For your safety and for protectio...

Страница 5: ...x 0 7 s max 9 s max 2 5 s from 0 to 80 of upper range limit from oo to 50 of upper range limit from 0 to 50 of upper range limit 3 Repair replacement of parts When opening the meter live parts may be exposed Therefore the meter must be disconnected from the measuring circuit prior to opening its case for repair or replacement of parts If repair cannot be avoided unless the meter is opened and live...

Страница 6: ...FF Press the ON OFF pushbutton 2 4 Function and range selection The function selector switch 6 is coupled with the Automatic terminal Blocking System ABS which allows access only to two correct sockets for each function Prior to switching to the mA or A functions or from the mA or A functions remove the test lead from the corresponding socket When the test leads are plugged in the terminal blockin...

Страница 7: ...section 4 1 4 3 Manual range section You can switch OFF autoranging and select the ranges manually according to the table on the following page Manual mode is switched OFF when pushbutton AUTO MAN is pressed 4 for approximately 1s when the function selector switch 6 is operated or when the meter is turned OFF and ON again When switching back to autoranging from 30 mV or 300 mV ranges 3 V range is ...

Страница 8: ...999 is shown by the right triangle 13 6 DATA hold facility The DATA function allows to automatically hold the measured values This is particularly useful for instance when connecting the probes to the measuring point requires full attention When the measured value is applied and the condition according to the table shown below is met the meter holds the measured value on the digital display and em...

Страница 9: ...djustment on the 30 mV measuring range Connect the test leads to the meter and join the free ends After having selected the measuring range brie y press the yellow multi function pushbutton 5 The meter acknowledges zero setting by a sound signal the LCD shows 00 00 1 digit and the decimal point ashes The displayed voltage at the instant the pushbutton is pressed is used as reference value max 200 ...

Страница 10: ...1 0 29 500 V DC AC eff rms sine 10 min 3 kHz 1200 V 30 kHz 300V 100kHz 30 V 500 V DC AC eff rms sine continu ously 10 min 7 Range 3 V U E 1 5 V eff rms 100 V eff rms 30 V UE 15 V eff rms 300 V eff rms 300V UE 150 V eff rms 1000 V eff rms 8 On the range 3V square wave signal positive on one side 5 15 V f const not 163 84 Hz or integral multiple 9 Without sensor 8 Voltagemeasurement Voltagemeasureme...

Страница 11: ...ed alternate switching takes place between DC and AC on the ZIEGLER Multi12S 15S or between DC and DC AC on the ZIEGLER Multi16S The change over is acknowledged by a sound signal The symbols DC and AC 11 are displayed as per selected function on the LCD When selecting a range with the function selector switch 6 the DC function is always set by default When pressing the yellow multi function pushbu...

Страница 12: ... 300 Digit Contin uously 0 75 3 10 Digit 1 0 5 10 Digit 1 0 5 10 D 1 0 2 0 25 2 1 0 5 10 Digit 1 0 2 1 0 5 10 Digit 1 0 2 1 5 2 10 Digit 1 5 2 10 Digit 1 5 2 10 Digit 1 5 2 10 Digit 1 5 2 10 Digit 0 36 A 5 0 36 A 5 6 6 6 6 0 36 A 5 0 5 3 4 0 5 3 0 4 1 0 4 1 0 4 1 0 6 1 2 0 1 0 25 1 0 5 5 10 D 0 5 2 0 5 5 10 D 0 5 2 1 0 5 10 D 1 0 2 1 5 4 10 D 1 5 4 10 D 1 75 4 10D 0 4 34 0 4 3 0 2 1 0 2 1 0 2 1 0 ...

Страница 13: ...tware MS WINDOWS 3 0 or higher Hardware IBM compatible PC with 1 MB main memory for the WINDOWS version A VGA or EGA monitor A hard disk with 2 MB free storage space A 3 5 disk drive for disks with 1 44 MB storage capacity A MICROSOFT compatible mouse If you wish to make hardcopy a printer supported by WINDOWS ...

Страница 14: ...ic load Caution The measuring ranges 3A and 16 A of ZIEGLER Multi13S are not protected by a fuse A blown fuse is signalled on the LCD the instant a measured quantity having a voltage of more than 4 V is applied to the corresponding connection sockets Then the digital display 9 shows the word FUSE After a fuse has blown eliminate the cause of the overload before using the meter again Replacement of...

Страница 15: ...unction selector switch 6 15 14 Duty cycle measurement ZIEGLER Multi 15S and 16S With duty cycle measurement we can determine the ratio of pulse duration to cycle time of recurring square wave signals Set the function selector switch 6 to V or V Connections are made in the same way as for voltage measurement See foot note 8 on page 21 Brie y press the yellow multi function pushbutton 5 twice The m...

Страница 16: ...Cll format for further processing For the use of ZIEGLER com 100 the following requirements must be met ON 13 Zero adjustment on the 30 Wmeasuring range When measuring small resistance values on the 30 W range you can eliminate the resistance of the leads and contact resistance by zero adjustment Connect the test leads to the multimeter and join the free ends Brie y press the yellow multi function...

Отзывы: