Each time the yellow multi-function pushbutton (5) is briey pressed,
the display changes between measured temperature and correction
value of the lead resistance.

We can exit the temperature measurement function

by pressing the yellow multi-function switch (5) longer, this is
conrmed by the two sound signals.
by changing the function selector switch.

Note:

For the lead resistance, the actual value measured on the digital
multimeter should be taken as correction value and not any specied
value.

16. Computer interface via RS 232C

The 

ZIEGLER

 

Multi 

 

S multimeters are tted with a serial RS-232C interface

for transmission of measured data to computer. The measured values
are optically transmitted through the case with infrared light to an interface
adapter which is attached to the multimeter. The measured data is passed
to the computer via a cable.

Switching the interface ON

When switching on the multimeter, press the "ON/OFF"  pushbutton
(2) and the " DATA--MIN/MAX" pushbutton (3) together.

With interface switched ON, automatic turn-OFF of the meter is
inactive. This is shown on the LCD (1) by ashing of the 

 (8) symbol.

The ’’DATA " function cannot be activated.

Interface packs as accessories

Interface adapter with memory :

By using this adapter it is possible to store the measured data without
interfacing the multimeter to the computer. This stored data can be
transferred later on to a computer. For establishing a powerful multimeter
system you can  connect upto ten multimeters ofine. Online connection
upto six multimeters to the computer is possible via memory adapter.

The interface packs include the adapters, the necessary connection
cables and the " ZIEGLER com 100 "  data acquisition and evaluation software
with operating instructions.

RISH com 100 software
The ZIEGLERcom 100

 software package corsists of a  WINDOWS version.

With ZIEGLER com 100, you can simultaneously access, store, display
and document the measured data of several 

ZIEGLER

 

Multi

 S multimeters.

The measured values are presented :

As digital display and analog indication similar to that of the
multimeter. (up to 4 multimeters)
In traces (XY and Yt) as on a 4-channel recorder.
In tabular form (data logger: upto 10 channels).

The measured data is stored in ASCll format for further processing.
For the use of ZIEGLER com 100, the following requirements must be met:

ON

13

Zero adjustment on the 30

measuring range

When measuring small resistance values on the 30 

W  range, you can

eliminate the resistance of the leads and contact resistance by zero
adjustment.

Connect the test leads to the multimeter and join the free ends.
Briey press the yellow multi-function pushbutton (5). The meter
acknowledges zero adjustment by a sound signal, the LCD shows
"00.00"(+1digit) and the decimal point ashes. The resistance measured
at the instant the pushbutton is pressed is used as reference value
(max.200 digits) It is automatically deducted from the values measured
thereafter. Zero adjustment can be cleared.
By pressing the yellow multifunction pushbutton (5) for a long time and
is acknowledges by two sound signals.
By switching the multimeter OFF.

11. Diode test and continuity test

Verify that the device under test is electrically dead. External voltages
would falsify the measured results!
Set the function selector switch (6) to "

"

connect the device under test as shown.

Forward direction and/or short circuit:
The multimeter displays the forward vol tage in Volts. As  long as the voltage
drop does not exceed the maximum display value of 1.999V, you can also
rest several series-connected elements or reference diodes with small
reference voltage.
Reverse direction or open circuit:
The multimeter indicates overrange "OL"

Note:

Resistors and semiconductor junction in parallel with the diode falsify the
measured results!

Diode test and continuity test with buzzer

With the "buzzer" function selected, the meter emits a continuous sound
signal on the range 0...approx. 1V.
To switch the buzzer ON:

Briey press the yellow multi-function pushbutton (5).
The multimeter acknowledges turn-ON with a sound signal. At the same

time, the symbol 

(18) appears on the LCD.

To switch the buzzer OFF

Briey press the yellow multi-function pushbutton (5) again.

 The multimeter acknowledges turn-OFF with a sound signal. The

symbol

(18) disappears from the LCD.

When selecting the function "Diode test and continuity test’’ with the function
selector switch (6), the buzzer is always switched OFF. Repeated brief
pressing of the multifunction pushbutton (5) alternately switches the buzzer
on and off. When pressing the push button for a long time, the buzzer is
always switched OFF, this is acknowledged by the buzzer
sounding twice.

Содержание MULTI 12S

Страница 1: ...Printed in India Subject to change without Notice Genius 04 05 2T User Manual ZIEGLER 12S 16S Analog Digital Multimeter Multi ...

Страница 2: ...10 A fused 1000 V max 7 10 0 20 30 ON OFF TRMS MAN MAX DATA MIN MAX AUTO MAN F mA 0 C A F W Hz V V V mA 0 C A 8 9 10 11 12 16 15 14 13 18 17 19 VA MIN MAX DC AC ON 8 8 8 8 10 0 20 30 MAN DATA KW MW kHz 0 C mF nF mm 5 19 Servicing When you need service please contact ...

Страница 3: ...measurement Rish multi 15S and 16S 14 14 Duty cycle measurement Rish multi 15S and 16S 15 15 Temperature measurement 15 16 Computer interface via RS 232C 16 17 Speci cations 18 18 Maintenance 26 18 1 Battery 26 18 2 Fuses 27 18 3 Case 27 19 Servicing 29 1 Liquid crystal display 2 ON OFF pushbutton 3 Pushbutton for data hold and MIN MAX storage functions 4 Pushbutton for manual range selection 5 Mu...

Страница 4: ...R Multi 12S 16S are manufactured and tested in compliance with the safety standard DIN VDE0411 and IEC 1010 1 DIN EN61010 1 VDE0411 1 In case of incorrect use or careless handling the safety of both user and multimeter is not assured For proper use and safe handling it is absolutely necessary to read and understand the operating instructions before using the meter For your safety and for protectio...

Страница 5: ...x 0 7 s max 9 s max 2 5 s from 0 to 80 of upper range limit from oo to 50 of upper range limit from 0 to 50 of upper range limit 3 Repair replacement of parts When opening the meter live parts may be exposed Therefore the meter must be disconnected from the measuring circuit prior to opening its case for repair or replacement of parts If repair cannot be avoided unless the meter is opened and live...

Страница 6: ...FF Press the ON OFF pushbutton 2 4 Function and range selection The function selector switch 6 is coupled with the Automatic terminal Blocking System ABS which allows access only to two correct sockets for each function Prior to switching to the mA or A functions or from the mA or A functions remove the test lead from the corresponding socket When the test leads are plugged in the terminal blockin...

Страница 7: ...section 4 1 4 3 Manual range section You can switch OFF autoranging and select the ranges manually according to the table on the following page Manual mode is switched OFF when pushbutton AUTO MAN is pressed 4 for approximately 1s when the function selector switch 6 is operated or when the meter is turned OFF and ON again When switching back to autoranging from 30 mV or 300 mV ranges 3 V range is ...

Страница 8: ...999 is shown by the right triangle 13 6 DATA hold facility The DATA function allows to automatically hold the measured values This is particularly useful for instance when connecting the probes to the measuring point requires full attention When the measured value is applied and the condition according to the table shown below is met the meter holds the measured value on the digital display and em...

Страница 9: ...djustment on the 30 mV measuring range Connect the test leads to the meter and join the free ends After having selected the measuring range brie y press the yellow multi function pushbutton 5 The meter acknowledges zero setting by a sound signal the LCD shows 00 00 1 digit and the decimal point ashes The displayed voltage at the instant the pushbutton is pressed is used as reference value max 200 ...

Страница 10: ...1 0 29 500 V DC AC eff rms sine 10 min 3 kHz 1200 V 30 kHz 300V 100kHz 30 V 500 V DC AC eff rms sine continu ously 10 min 7 Range 3 V U E 1 5 V eff rms 100 V eff rms 30 V UE 15 V eff rms 300 V eff rms 300V UE 150 V eff rms 1000 V eff rms 8 On the range 3V square wave signal positive on one side 5 15 V f const not 163 84 Hz or integral multiple 9 Without sensor 8 Voltagemeasurement Voltagemeasureme...

Страница 11: ...ed alternate switching takes place between DC and AC on the ZIEGLER Multi12S 15S or between DC and DC AC on the ZIEGLER Multi16S The change over is acknowledged by a sound signal The symbols DC and AC 11 are displayed as per selected function on the LCD When selecting a range with the function selector switch 6 the DC function is always set by default When pressing the yellow multi function pushbu...

Страница 12: ... 300 Digit Contin uously 0 75 3 10 Digit 1 0 5 10 Digit 1 0 5 10 D 1 0 2 0 25 2 1 0 5 10 Digit 1 0 2 1 0 5 10 Digit 1 0 2 1 5 2 10 Digit 1 5 2 10 Digit 1 5 2 10 Digit 1 5 2 10 Digit 1 5 2 10 Digit 0 36 A 5 0 36 A 5 6 6 6 6 0 36 A 5 0 5 3 4 0 5 3 0 4 1 0 4 1 0 4 1 0 6 1 2 0 1 0 25 1 0 5 5 10 D 0 5 2 0 5 5 10 D 0 5 2 1 0 5 10 D 1 0 2 1 5 4 10 D 1 5 4 10 D 1 75 4 10D 0 4 34 0 4 3 0 2 1 0 2 1 0 2 1 0 ...

Страница 13: ...tware MS WINDOWS 3 0 or higher Hardware IBM compatible PC with 1 MB main memory for the WINDOWS version A VGA or EGA monitor A hard disk with 2 MB free storage space A 3 5 disk drive for disks with 1 44 MB storage capacity A MICROSOFT compatible mouse If you wish to make hardcopy a printer supported by WINDOWS ...

Страница 14: ...ic load Caution The measuring ranges 3A and 16 A of ZIEGLER Multi13S are not protected by a fuse A blown fuse is signalled on the LCD the instant a measured quantity having a voltage of more than 4 V is applied to the corresponding connection sockets Then the digital display 9 shows the word FUSE After a fuse has blown eliminate the cause of the overload before using the meter again Replacement of...

Страница 15: ...unction selector switch 6 15 14 Duty cycle measurement ZIEGLER Multi 15S and 16S With duty cycle measurement we can determine the ratio of pulse duration to cycle time of recurring square wave signals Set the function selector switch 6 to V or V Connections are made in the same way as for voltage measurement See foot note 8 on page 21 Brie y press the yellow multi function pushbutton 5 twice The m...

Страница 16: ...Cll format for further processing For the use of ZIEGLER com 100 the following requirements must be met ON 13 Zero adjustment on the 30 Wmeasuring range When measuring small resistance values on the 30 W range you can eliminate the resistance of the leads and contact resistance by zero adjustment Connect the test leads to the multimeter and join the free ends Brie y press the yellow multi function...

Отзывы: