LAP-F1 Logic Analyzer
|
User Guide v1.6
| www.zeroplus.com.tw
69
by the LAP-F1 as Low or 0, and voltages above are defined as
High or 1.
Table 4:17 Trigger Level dialog box description
4.17.2.
DSO Connection
A DSO Connection can be set up when users want to import and display a DSO signal
in the ZP-Logic software. This can be useful since the LAP-F1 does not have the
ability to capture analog signals. The supported DSO models are listed in Table 4:19.
Two operation modes are possible; the connection can be set up with LAP-F1 as
master or slave depending on which instrument the user wants to provide the trigger
signal. The two modes are described in continuation. The connection setup follows in
chapters 4.17.2.1 and 4.17.2.2.
4.17.2.1.
LAP-F1 as Master
When the LAP-F1 is the master the DSO is the slave. In this mode, the Trigger Out of
the LAP-F1 connects with the Trigger In of the DSO. When the trigger event occurs,
the LAP-F1 sends a trigger signal to the DSO which, upon receiving the signal, starts
to capture data. See the complete connection diagram in Figure 4-22.
Figure 4-22 Connection diagram with LAP-F1 as master