LAP-F1 Logic Analyzer
|
User Guide v1.6
| www.zeroplus.com.tw
70
4.17.2.2.
LAP-F1 as Slave
When the LAP-F1 is the slave the DSO is the master. In this mode, the Trigger Out of
the DSO connects with the Trigger In of the LAP-F1. When the trigger event occurs,
the DSO sends a trigger signal to the LAP-F1 which, upon receiving the signal, starts
to store data. See the complete connection diagram in Figure 4-23.
Users can try to connect the DSO Trigger Out to any regular channel of the LAP-F1 if
the BNC connector is occupied by another instrument.
Figure 4-23 Connection diagram with LAP-F1 as slave
4.17.2.3.
Settings
Up to 4 analog signals can be shown; see the settings dialog box in Figure 4-24.