ERROR TITLE
ERROR MESSAGE
CAUSE / ACTIONS
Tablet Not Plugged in
Confirm the Tablet is plugged in�
Tablet is not plugged into power, power
adapter is not connected properly or
electrical outlet is not functioning�
•
Confirm power adapter is correctly
plugged into the Tablet and an
electrical outlet
•
Confirm the electrical outlet is
functioning
•
Leave the Tablet plugged in for the
rest of the night to ensure your study
continues
Nasal Cannula Not
Connected
Confirm the Nasal Cannula is
connected�
Nasal Cannula is not connected to
Recorder, or prongs of Nasal Cannula are
not positioned correctly in your nose, or
your nose is plugged/congested�
•
Disconnect/Reconnect Nasal Cannula
to Recorder
•
Ensure prongs of Nasal Cannula are
properly positioned in each nostril
•
Check the length of the Nasal Cannula
for kinks
•
Blow your nose
•
If error continues, stop the study and
contact your Healthcare Provider
Oximeter Not
Connected
Confirm the Oximeter is
connected
Pulse Oximeter is not connected to
Recorder, or not positioned properly on
your finger�
•
Disconnect/Reconnect Pulse Oximeter
to Recorder
•
Ensure sensor is properly positioned on
your finger
•
If error continues, stop the study and
contact your Healthcare Provider