Warnings
1.
Federal Law restricts this device to sale by or on the order of a licensed practitioner� Prescription use
only� To be used only for prescribed purposes�
2.
Read all manuals and labels prior to use� Ensure that you have been trained prior to use�
3.
Use of the device during a Sleep Theragnostic may cause:
•
Temporary bite changes
•
Gum or tooth discomfort
•
Jaw discomfort
•
Jaw joint discomfort
•
Increased salivation
•
Dry mouth
4.
Use of the MATRx plus may cause disruption in sleep from the use of the sensors and Mandibular
Positioner�
5.
Stop use of the device during a Sleep Theragnostic if you experience any of the following:
•
Nausea or vomiting
•
Jaw or mouth pain
•
Adverse reaction to any system component
6.
Electric shock hazard� Do not remove or disassemble any covers, enclosures or parts of system
components� The system is not user serviceable� Do not use the device near liquids� Contact Zephyr
for service or repair inquiries�
7.
Do not charge the Recorder while it is connected to you�
8.
This device needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided (see Manufacturer Declaration for Emissions)�
9.
Portable and mobile RF communications equipment such as wireless home network devices, mobile
phones, cordless telephones and their base stations, walkie-talkies can affect this device and should
be kept at least a distance 30 cm away from the equipment�
10.
The device cannot be used during defibrillation�
11.
Do not use extension cords or adapters of any type for device charging� The power cord and plug must
be intact and undamaged� Use only the chargers provided�
12.
Fire, explosion or severe burn hazard� Do not crush, disassemble, heat above 100C (212F), incinerate
or expose the device to water�
13.
Explosion hazard� Do not use the Pulse Oximeter in the presence of flammable anesthetics or other
flammable substances in combination with air, oxygen-enriched environments, or nitrous oxide�
14.
Do not connect to an electrical outlet controlled by a wall switch or dimmer�
15.
Do not place the device or external power supply in any position that might cause it to fall on you� Do
not lift the device by the power supply cord or patient cable; use only the body of the device�