4 Installation | 4.4 Connecting the Microscope to the Power Supply
ZEISS
4.4 Connecting the Microscope to the Power Supply
4.4.1 Connecting the Microscope to the Mains Power Supply
All microscopes equipped with Fixed-Köhler stands or Full-Köhler stands can be connected directly
to the mains power supply.
Prerequisite
ü
The microscope is installed completely with all components to be used.
Procedure
1. Plug the power cable
3
into the mains socket
2
of the microscope.
2
3
1
4
5
Fig. 30: Connecting the microscope to the mains power supply
2. Make sure the
DC 5V
switch
1
is in the
OUT
position.
à
The
DC 5V
switch is only available on Fixed-Köhler stands.
3. Plug the other end
5
of the power cable into the mains power supply
4
.
4.4.2 Connecting the Microscope to the Battery Supply Unit
All microscopes equipped with Fixed-Köhler stands can be connected directly to the mains power
supply.
Prerequisite
ü
The microscope is installed completely with all components to be used.
ü
Rechargeable batteries are inserted in the battery supply unit.
ü
The batteries are charged.
Procedure
1. Plug the connector
3
of the battery supply unit into the connection port
2
of the
stand.
2
3
1
4
5
Fig. 31: Connecting the microscope to the battery supply unit
2. Make sure the
DC 5V
switch
1
is in the
OUT
position.