4 Installation | 4.3 Installing Additional Components
ZEISS
4.3.10 Installing/Removing the Condenser
The mirror serves to illuminate the specimen if no power outlet is available.
Parts and Tools
Hex Key
Prerequisite
ü
Full-Köhler stand
Procedure
1. Turn coarse focusing drive to move the stage
carrier to the upper stop position.
2. Lower the condenser carrier as far as it will go,
using the knurled screw for vertical adjustment.
3. Loosen the clamping screw of the condenser, so that the condenser can be removed front-
ward.
à
Use an Allen wrench if the clamping screw is
an Allen screw.
4. Remove the installed condenser, e.g. Abbe condenser.
5. Insert the condenser to be installed, e.g. turret condenser, with its dovetail ring mount into
the condenser carrier.
6. Fix it by tightening the clamping screw.
54
Instruction Manual ZEISS Primostar 3 | en-US | Rev. 1 | 415501-7011-111_en