162
Preparations
G-30-1530-en
OPMI® VISU 210 on S8, S81, S88 Suspension Systems
Issue 1.0
Printed on 19. 04. 2004
Adjusting the tilt angle
Using knob (1), you can position the surgical microscope in a range from
+180° to -180° (+ in the direction of the surgeon and - in the opposite di-
rection). The +90° setting is ideal for surgery on patients in a seated po-
sition or lying on their side.
Caution:
Do not tilt the main microscope / -180°, as this could damage
the microscope cable or the light guide.
•
Turn knob (1) until the surgical microscope is in the viewing position
required.
After the viewing angle has been set, the surgical microscope remains in
this position. The gear drive is self-locking.
Warning!
–
To prevent the assistant's microscope from moving downward of its
own accord when the main microscope is being tilted, the assistant's
microscope must be adjusted and locked in position using screw (2)
before surgery.
–
Check that the assistant's microscope is firmly seated.
–
Do not tilt the main microscope / - 180°, as this could dam-
age the microscope cable or the light guide.
•
Using locking screw (2), secure the assistant's microscope in the se-
lected working position (on the left or right of the main microscope).
•
Turn knob (1) until the surgical microscope is in the viewing position
required.
After the viewing angle has been set, the surgical microscope remains in
this position. The gear drive is self-locking.
MISU241
Содержание OPMI VISU 210
Страница 29: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Safety 29 MISU241...
Страница 37: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Safety 37 1 MISU241...
Страница 111: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Description 111 2 1 MISU241...
Страница 145: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Preparations 145 MISU241...
Страница 147: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Preparations 147 1 2 MISU241...
Страница 149: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Preparations 149 1 MISU241...
Страница 151: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Preparations 151 1 2 MISU241...
Страница 153: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Preparations 153 1 MISU241...
Страница 157: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Preparations 157 1 2 MISU241...
Страница 159: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Preparations 159 1 MISU241...
Страница 160: ...160 Preparations G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 MISU241...
Страница 163: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Preparations 163 1 2 MISU241...
Страница 165: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Preparations 165 1 2 MISU241...
Страница 173: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Operation 173 MISU241...
Страница 175: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Operation 175 1 2 MISU241...
Страница 199: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 Operation 199 1 MISU241...
Страница 202: ...202 Operation G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 MISU241...
Страница 251: ...G 30 1530 en OPMI VISU 210 on S8 S81 S88 Suspension Systems Issue 1 0 Printed on 19 04 2004 251 MISU241...