Roll-over procedure
Use the "Roll-over procedure" to select the required function in the default
sequence of the above table.
Every time you press the relevant button "
∇
" you advance clockwise.
Every time you press the relevant button "
∆
" you advance counterclock-
wise.
→
XYZ-RES
→
FOC-MEM
→
XY-INV
→
↑
↓
←
XY-RES
←
AUX
←
PHOTO
←
SDI
←
If you have assigned FOC-MEM to key C or D of the foot control panel,
you can determine by the length of time for which you press the relevant
key of the foot control panel during operation whether a positioning run is
to be triggered (press the key for less than 2 seconds) or whether the cur-
rent position is to be saved (press the key for more than 2 seconds).
Caution:
Be extremely careful when changing these settings. You should change
settings only under your own user ID. Notify all users of any changes, or
make sure that each user only works under his own user ID.
Saving parameter settings
As soon as you have entered a parameter setting, it is saved under the
current user ID.
Description of the assignable functions:
Path: To return to the basic mode, press the "MODE" key in configuration
mode 1 or do not press any key for 20 seconds.
To switch to configuration mode 2, simultaneously press the
"MODE" and "STORE" keys.
XY-RES
Recenters the X-Y coupling and resets the focus to its in-
itial position in the focusing range.
XYZ-RES
Recenters the X-Y coupling, resets the focus to its initial
position in the focusing range, and sets the zoom to a po-
sition previously stored using the STORE key.
XY-INV
Inverts the direction of movement of the X-Y coupling. This
function is useful when you are working with an image re-
versal system in vitreo-retinal surgery. You can now invert
the direction of movement of the X-Y coupling by tipping
on one of the two keys (C or D) of the foot control panel.
SDI
The optical system of the Stereo Diagonal Inverter (SDI)
2E from Oculus is moved into and out of the beam path of
the surgical microscope.
Содержание OPMI VISU 160
Страница 31: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Safety 31...
Страница 33: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Safety 33 1 1 2 3...
Страница 37: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Safety 37 4 5 3 2 1...
Страница 41: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Safety 41 1...
Страница 48: ...48 Safety G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006...
Страница 51: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Description 51...
Страница 56: ...56 Description G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006...
Страница 101: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Description 101 2 1 4 5 3...
Страница 105: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Description 105 1 3 2 4 5...
Страница 107: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Description 107 3 1 2 2 2 4...
Страница 111: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Description 111 3 4 5 1 2...
Страница 113: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Description 113 1 2 3 4 1 5...
Страница 115: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Description 115 3 2 1...
Страница 117: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Description 117 3 2 1...
Страница 119: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Description 119 3 2 1...
Страница 121: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Description 121 2 1 3 1...
Страница 131: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Preparations 131 1 2 3 4...
Страница 137: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Preparations 137 2 3 4 5 6 1...
Страница 139: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Preparations 139...
Страница 141: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Preparations 141 1 2...
Страница 143: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Preparations 143 1...
Страница 150: ...150 Preparations G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006...
Страница 157: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Operation 157 1 2...
Страница 177: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Operation 177 7 6 1 2 3 4 5...
Страница 181: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Operation 181 1...
Страница 184: ...184 Operation G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006...
Страница 215: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 Technical data 215 270...
Страница 233: ...G 30 1529 en OPMI VISU 160 on S8 S81 S88 Suspension Systems Issue 5 0 Printed on 09 10 2006 233...