SET-UP 

Mechanical stage S mot. CAN 

Carl Zeiss 

General information 

Measuring stage S mot. CAN 

 
 

 

14 

435465-9000-701 

08/2012 

3

 

SET-UP 

3.1

 

General information 

 

Read the 

Notes on instrument safety

 carefully prior to set-up and operation (see Section 1.2, 

page 7). 

The mechanical stage / measuring stage S mot. CAN is normally delivered together with the required 
tools and optional accessories in several packages.  

 

Remove all components from the packaging and check that all components described on the delivery 

note are present. 

 

Remove all shipping braces (adhesive tape or similar). 

 

Keep the original packaging for storage or for returning the instrument to the manufacturer, or 
dispose of it properly.  

 
 
 

3.2

 

Mounting the mechanical / 

measuring stage S mot. CAN 

 

 

In the event that a microscope system 

is not yet available or assembled, the 
microscope must initially be installed 

and put into operation according to 

the separate operating  manual. The 
transillumination top 450 mot.  may 

already be mounted. 

 

 

Switch off the microscope and illumination. 

 

Remove the shipping screw (Fig. 2/

2

) on the 

underside of the respective stage and keep it for 

later use. 

 

Only the measuring stage S mot. CAN  
has a trigger socket (Fig. 2/

1

) on its 

underside. 

 

 

Fig. 2 

Removing the transport locking screw  

 
 

Содержание Measuring stage S mot.

Страница 1: ...Operating Manual Mechanical Measuring stage S mot for Axio Zoom and SteREO Discovery...

Страница 2: ......

Страница 3: ...d a high quality product from Zeiss Before using the instrument the first time please read this operating manual in order to maintain the high quality of the instrument and ensure reliable work with i...

Страница 4: ...anual is not subject to updating or revision Unless explicitly authorized dissemination or duplication of this document and commercial exploitation or communication of its contents are not permitted P...

Страница 5: ...1 General information 14 3 2 Mounting the mechanical measuring stage S mot CAN 14 3 2 1 Mounting the mechanical measuring stage S mot CAN on base 300 for epi illumination 15 3 2 2 Mounting the mechan...

Страница 6: ...magnetic Compatibility and 2006 42 EC Machinery Directive with the appropriate controller EMS 3 The operating manual contains all information and warnings the operator must comply with The following w...

Страница 7: ...itched off and secured against inadvertent operation Please contact a Zeiss service agency or the Carl Zeiss Microscopy Service to repair the instrument Dust and dirt may impair the instrument s perfo...

Страница 8: ...P 3 is at the zero position and not swiveled out and that the foot pedal is clear Automatic travel of the motorized tables can be interrupted by pressing the STOP button on the focusing drive or the S...

Страница 9: ...wear wearing parts in particular and improper use The instrument manufacturer shall not be liable for damage caused by faulty operation negligence or any other tampering with the instrument particular...

Страница 10: ...es When using the mechanical measuring stage S mot CAN the separate operating manuals for the Axio Zoom V16 or SteREO Discovery and insofar as they are used the Transillumination top 450 mot SYCOP 3 s...

Страница 11: ...010 000 EMS 3 Required for operating the stage USB 2 0 cable to workstation with software Country specific power cable SYCOP 3 system control panel 435611 9010 000 SYCOP 3 for stage control Cable MDR...

Страница 12: ...Objective E Nosepiece F Intermediate tube Fluar Illuminator Z mot G Camera adapter with microscope camera AxioCam H Reflected light illumination slit ring illuminator with cold light source CL 9000 L...

Страница 13: ...kaging Permissible ambient temperature 5 to 40 C Permissible relative humidity max 75 at 35 C non condensing Transport in packaging Permissible ambient temperature 40 to 70 C Operation Permissible amb...

Страница 14: ...l shipping braces adhesive tape or similar Keep the original packaging for storage or for returning the instrument to the manufacturer or dispose of it properly 3 2 Mounting the mechanical measuring s...

Страница 15: ...o this push the plate Fig 4 3 into the stage Fig 4 1 with the left front corner slightly inclined allow it to slide under slight pressure against the springs Fig 4 4 until it is completely inside the...

Страница 16: ...into the stand base 450 and secure with the three captive screws Fig 5 6 Detach the insert plate Fig 5 3 from the stand base 450 Fig 5 5 Set the mechanical measuring stage Fig 6 3 on the insert plate...

Страница 17: ...w it hand tight with the 4 captive screws Set the mechanical measuring stage Fig 7 5 on the transillumination base 300 stage drive pointing to the left align to the rear edge and secure with the three...

Страница 18: ...4 Remove the three short hexagon socket screws Fig 8 5 on the insert plate and keep for possible later use Remove assembly aid d 120 mm Fig 8 2 again Set the mechanical measuring stage Fig 8 7 on the...

Страница 19: ...tension cable HD15 Fig 9 3 via the upper connecting socket Fig 9 1 Connect SOCP 3 Fig 9 5 to EMS 3 In addition to or as an alternative to SYCOP 3 the mobile coaxial drive XY CAN can be connected using...

Страница 20: ...sed while moving the joystick accordingly If SYCOP 3 is not available on the microscope system the mobile coaxial drive XY CAN can be optionally used for external control of the stage see Section 4 2...

Страница 21: ...p Mode Stage Calibrate Calibration is necessary for use of the SYCOP 3 memory function It is recommended that this is carried out immediately upon prompting The settings in the Memory menu are thus av...

Страница 22: ...positioned left and right Fig 11 2 or 3 Saving the current XY position by holding down the respective button Going to the saved XY position by briefly pressing the respective button The XY positions a...

Отзывы: