2  LSM 710 – Quick Guide                                                MFPL BioOptics - Light Microscopy 

 

 
Shutdown 

1.

 

Switch the Argon laser from 'run' to 'idle' (see image below, #1). 

2.

 

Turn the key on the Argon laser power supply 90° counterclockwise from position 1 to 0 (see 
image below, #2). 

3.

 

Close ZEN. In the laser tool window popping up, turn off all lasers (see image below, #3). Turn off 
the computer. 

4.

 

Turn off the X-Cite lamp (see image below, #4). 

5.

 

Push the 'Load position' button on the touch panel (on the microscope table) to lower the 
objective (see below, #5). Remove the sample and clean oil objectives with lens cleaning paper. 

6.

 

Wait until the fan of the Argon laser has stopped. Then turn off the Systems/PC, Components and 
Main Switches (see below, #6). 

7.

 

Put the dust cover on the microscope. 

 

Leaving the system for the next user

 

1.

 

Switch the Argon laser to 'idle' mode (see image above, #1). 

2.

 

Push the 'Load position' button on the touch panel (see image above, #5). Remove the  sample 
and clean oil objectives with lens cleaning paper. 

3.

 

In ZEN, close your images. If you have used the 561 and/or 633 nm laser, leave them running. 
Leave ZEN running. 

4.

 

Disconnect from servers: 
HIVE: click on the HIVE icon. In the dialog popping up, select 'Disconnect' 
VBC share: click on the 'Logout' icon 

 