Zeiss LSM 710
3 LSM 710 – Quick Guide MFPL BioOptics - Light Microscopy
Troubleshooting
ZEN hangs during startup
1.
In the login screen, click on the triangle left to 'Boot Status' (see the image below, #1).
2.
Check the messages in the expanding scroll pane (see the image below, #2).
3.
If there's a message "No connection to realtime controller", open the the door of the RTC
(black box left to the microscope) , press the Reset Button (see image below, #3) and wait
until ZEN resumes.