B-0705-1 01/1992 (Digitized 07/2005)
2-6
(1)
Setting up and connecting the control panels
Control panel BA (2-10/3) and control panel LO (2-10/1) to the left or to the right order from the stand (2-10/2)
connect to the power supply (2-10/4) and secure with screws (2-10/5).
NOTE
The plug connections between the operator panels and the stand may not
Are released or produced voltage.
(2)
Connect the mains voltage
CAUTION
Check that the on the rear panel
value printed on the network-
voltage corresponds to!
• Power cord (2-11/4) in device connection (2-11/2)
plug it in and connect to the network
produce.
• Switch, network switch (2-11/1) in
Position "1". The operational readiness is
indicated by the LED on the control panels.
NOTE
In case of failure both fuses
control: F1, F2:
2AT (100-127 V), 1AT (200 - 240 V)
For this purpose, the fuse holder (2-11/3) from
take the case by the
two spring tabs simultaneously in
Direction of the arrow pressed.
Figure 2-11 Connecting the mains voltage
View from behind
Figure 2-10
Page 40
2.2
Operation
2.2.1
Initial start-up
The first time the microscope JENA PLAN The following work is performed:
Содержание JENA PLAN HD 9x12
Страница 2: ...Original text Page 1 Contribute a better translation Operation instructions JENA PLAN...
Страница 9: ...B 0705 1 01 1992 Digitized 07 2005 VI OVERALL VIEW Page 9...
Страница 14: ...B 0705 1 01 1992 Digitized 07 2005 1 2 Figure 1 1 microscope equipment JENA PLAN Page 14...
Страница 15: ...B 0705 1 01 1992 Digitized 07 2005 1 3 Page 15...
Страница 17: ...B 0705 1 01 1992 Digitized 07 2005 1 5 Page 17...
Страница 18: ...B 0705 1 01 1992 Digitized 07 2005 1 6 Page 18...
Страница 19: ...B 0705 1 01 1992 Digitized 07 2005 1 7 Page 19...
Страница 20: ...B 0705 1 01 1992 Digitized 07 2005 1 8 Page 20...
Страница 33: ...Page 33 B 0705 1 01 1992 Digitized 07 2005 1 22...
Страница 67: ...B 0705 1 01 1992 Digitized 07 2005 2 35 Page 69 3 Care Maintenance Service 3 1 Care...
Страница 75: ...B 0705 1 01 1992 Digitized 07 2005 A 4...