Set-up
Fluar Illuminator Z mot.
Connecting the Fluar Illuminator Z mot.
Carl Zeiss
05/2012
435180-9060-701
17
Removal of the manual stop slider
•
Insert the ball-head screwdriver (SW 3 mm) into
the hole (Fig. 7/
2
) and push its indentation
upwards into the groove.
•
Remove the stop slider with the ball-head
screwdriver.
Removal of the motorized stop slider
•
Insert the ball-head screwdriver into the hole
(Fig. 7/
1
) press against the spring element until
the stop slider projects slightly.
•
Remove the stop slider by hand.
•
If the stop slider is not required, remove the
slotted cover and replace it with the blind cover.
For special fluorescence applications (producing optical cross-sections through the specimen)
the ApoTome.2 system can be coupled with the Fluar Illuminator Z mot.
For this purpose the diaphragm shaft is provided with a special cover which permits mounting
of the ApoTome.2 slider. This cover is supplied with the ApoTome.2 system if required.
3.6
Connecting the Fluar Illuminator Z mot.
Connecting to a microscope system with coarse/fine drive
•
Connect the CAN-RJ45 socket on the reverse side of the Fluar Illuminator Z mot. to the mains outlet
with a plug-in power unit.
Connecting to microscope system with focusing motor or SYCOP 3 / EMS 3
•
Connect the Fluar Illuminator Z mot. to an empty CAN socket on the rear side of the focusing motor
using a 0.5 m CAN-RJ45 cable or to the EMS 3 controller using a 1.0 m CAN-RJ45 cable.
Fig. 7
Removing the stop slider