OPERATION
Carl Zeiss
Experiment Procedure
DirectFRAP
48
423635-9001-701
05/2014
Application
Laser lines for
manipulation
(m) and imaging
in combination
systems (i)
Filter sets and filters for finding cells
Filter sets and filters for the manipulation
experiment
Combination
a)
with
Cell Observer SD
"Stand alone" DirectFRAP
Combination
a)
with TIRF 3
"Stand alone" DirectFRAP
Combination
a)
with TIRF 3
Reflector
turret
Excitation
filter wheel
Reflector turret
Reflector
turret
Excitation
filter wheel
(s)
Reflector turret
Beam
combiner
80/20
Fast beam
path
switch
Bleaching of GFP
or emerald
488m/i
FS 38 or FS
74
Empty
position
FS 74
FS 74
Empty
position
FS 74
No additional component
required
Conversion of
Dendra or Eos
405m
488m/i, 561i
FS 38
Empty
position
FS 76
FS 82
b)
Additional
excitation
filter from FS
82
FS 76
Reversible on/off
switching of
Dronpa
405 (on)m,
488 (off)m/i
FS 83 (on);
FS 38, 52 or
74 (off)
Additional
excitation
filter from FS
83 (on)
FS 76 or 83 (on);
FS 52 or 74 (off)
FS 83 (on);
FS 52 or
74 (off)
Additional
excitation
filter from FS
83 (on)
FS 76 or 83 (on);
FS 52 or 74 (off)
a)
For combinations, a 100 % mirror inserted in the DirectFRAP is recommended for maximum dynamics to increase efficiency.
b)
The manipulation of Dendra with the laser line 488 is not possible in the "stand alone" DirectFRAP system.
c)
FS 18 cannot be used for PA-GFP. For finding cells, a transmitted-light contrast is required.
d)
No laser is available for imaging of the initial form of PS-CFP or Phamret.