Setting up the microscope 

 

Changing between direct observation and laser scanning mode 

The 

Ocular

 and 

Acquisition

 buttons switch between the use of the LSM and the microscope: 

 

!

 

Click  on  the 

Ocular

  button  to  open  the  controls  for  the 

microscope  beam  path  and  for  direct  observation  via  the 
eyepieces of the binocular tube, lasers are blocked. 

!

 

To  set  the  hardware  in  position  for  using  the  microscope, 
click 

Online

 if not yet active. 

!

 

To close the light shutters on the microscope click 

Offline

 

!

 

Click  on  the 

Acquisition

  button  to  move  back  to  the 

LSM system. 

 
 

Setting up the microscope and storing 
settings 

Click  on  the 

Ocular

  tab  for  direct  observation; 

press  the 

Online

  button  for  your  actions  to  take 

effect  immediately.  Then  open  the 

Ocular

  tool  to 

configure the components of your microscope like 
filters, shutters or objectives (Fig. 11). 

Selecting an objective 

!

 

Open the graphical pop-up menu by clicking on 
the 

Objective

  symbol  and  select  the  objective 

lens for your experiment (Fig. 11). 

!

 

The  chosen  objective  lens  will  automatically 
move into the beam path. 

 

Focusing the microscope for transmitted light 

!

 

Open the graphical pop-up menu by clicking on 
the 

Transmitted Light

 icon (Fig. 12). 

!

 

Click on the 

On 

button. Set the intensity of the 

Halogen lamp using the slider. 

!

 

Clicking outside the pop-up control closes it. 

!

 

Place specimen on microscope stage. The cover 
slip  must  be  facing  the  objective  lens.  Re-
member the immersion medium if the objective 
chosen requires it! 

 

Fig. 11 

Microscope Control window, e.g.: 
Axio Imager.Z2 

 

 

 

 

Содержание ConfoCor 3

Страница 1: ...011234 5 66...

Страница 2: ......

Страница 3: ...es from his samples This Quick Guide does NOT replace the detailed information available in the full user manual or in the manual of the respective microscopes Axio Imager Axio Observer Axio Examiner...

Страница 4: ...or reflected light illumination via the power supply as described in the respective operating manual Switching on the Ar ML Laser If the Ar ML laser is required switch it on via the toggle switch Fig...

Страница 5: ...tes the entire software package for new image acquisition and analysis The Image Processing mode ignores all hardware and activates only data handling and image processing functionality for already ac...

Страница 6: ...4 04 2009 Fig 4 ZEN Main Application Window after Startup with empty image container Fig 5 ZEN Main Application Window after Startup with several images loaded...

Страница 7: ...the image According to the chosen view tab the required view controls appear in View Control Tabs below each image File management and data handling tools are found in the Right Tool Area see Fig 4 a...

Страница 8: ...your personal preferences Fig 7 Setting up conventional confocal software for a specific experiment can take a long time and is often tedious to repeat With ZEN these adjustments have to be done only...

Страница 9: ...written by a new scan Multi dimensional scans or saved images will never be over written and a new scan will then automatically create a new image document Acquired data is not automatically saved to...

Страница 10: ...File Menu The ZEN File Browser can be used like the WINDOWS program file browser Images can be opened by double click and image acquisition parameters are displayed with the thumbnails Fig 9 For more...

Страница 11: ...the lasers are turned on automatically The Laser Life Extender function of the software shuts all lasers off if ZEN is not used for more than 15 minutes To manually switch lasers on or off Click the...

Страница 12: ...press the Online button for your actions to take effect immediately Then open the Ocular tool to configure the components of your microscope like filters shutters or objectives Fig 11 Selecting an obj...

Страница 13: ...icroscope settings Microscope settings can be stored as configurations Fig 13 by typing a config name in the pull down selector and pressing the save button Fast restoration of a saved config is achie...

Страница 14: ...rrow in the dye list and simply choose the dye s you want to use in your experiment from the list dialogue In this dialogue the dyes can be also searched by typing the name in the search field Fig 14...

Страница 15: ...n If the option Linear Unmixing is selected the system is set in the lambda mode automatically Pressing the Auto Exposure button will then optimize the settings of the Gain Master and offset for the g...

Страница 16: ...uble and triple labeling line by line or frame by frame Advantage Only one detector and one laser are switched on at any one time This reduces cross talk Disadvantage slower image acquisition Open the...

Страница 17: ...box of the selected channel Ch 1 4 monitor diode ChM QUASAR detectors ChS1 8 transmission ChD for the scanning procedure and assigning a color to the channel Select the appropriate filters and activat...

Страница 18: ...and channels Frame Tracks are switched during scanning frame by frame The following settings can be changed between tracks Laser line and intensity all filters and beam splitters the channels incl se...

Страница 19: ...the Acquisition Mode tool Fig 20 to adjust the scan speed A higher speed with averaging results in the best signal to noise ratio Scan speed 8 usually produces good results Use speed 6 or 7 for superi...

Страница 20: ...mode in the Acquisition Mode tool Select the number of lines or frames to average Adjusting pinhole size Select the Channels tool in the Left Tool Area Set the Pinhole size to 1 AU Airy unit for best...

Страница 21: ...pre adjustment of detector gain and offset Select Live for continuous fast scanning useful for finding and changing the focus Select Continuous for continuous scanning with the selected scan speed Se...

Страница 22: ...1 Airy Unit Fig 25 Set the Gain Master high When the image is saturated reduce AOTF transmission in the Laser control section of the Channels Tool Fig 25 using the slider to reduce the intensity of th...

Страница 23: ...Stack Then focus on the lower specimen area where the recording of the Z Stack is to end Click on the Set Last button to set this lower position Click on the button to set number of slices to match th...

Страница 24: ...r experiment Click on the Save button If you close an image which has not been saved a pop up window will ask you if you want to save it Choosing yes will lead you to the WINDOWS Save As window To exp...

Страница 25: ...itch off the Ar ML laser with first the idle run switch Fig 2 3 and second the key switch Fig 2 1 then wait until the fan of the Argon laser has switched off Don t switch off the main power yet On the...

Отзывы: