Maintenance and care
000000-2121-813-GA-GB-050416
40
General information
Do not use acetone and acetone-based cleaning agents to clean the device,
as they could damage the surfaces.
Use spray or wipe disinfectants for cleaning and disinfecting the instrument
casing, chinrest and forehead strap.
Heavily soiled painted surfaces should be cleaned with a cloth moistened
with a weak soap solution.
When cleaning the monitor, use only denatured alcohol or commercially
available cleaning cloths. Use commercially available cleaning cloths for the
mouse and keyboard.
To clean the printer, please observe the information given in the
manufacturer’s operating instructions.
Replacing the paper pads
•
Remove the two pins for securing the paper pads (6, Fig. 4) on the
headrest.
•
Place the paper pads on the chin rest (5, Fig. 4) of the headrest.
•
Re-insert the two holding pins through the holes in the paper pads and
into the holes provided in the headrest.
•
Tear off the top paper liner after each treatment.
Cleaning the front lens
The cleanliness of the front lens has a crucial influence on the quality of the
captured images.
Please ensure that the front lens is protected at all times from contamination
due to skin contact, body fluids and dust. Use the protective cover supplied
for the purpose.
Observe the following instructions when using the CIRRUS photo and for
cleaning the front lens:
•
Pull the CIRRUS photo towards the operator when not in use and lock it
to ensure the greatest possible clearance between patient and
instrument.
•
Do not remove the protective cover from the lens until all preparations
for capturing images have been completed and replace it immediately
after completing the examination.
•
Never remove the lens from its casing for cleaning.
Содержание CIRRUS photo 600
Страница 1: ...CIRRUS photo CIRRUS photo 600 and CIRRUS photo 800 Documentation set...
Страница 4: ......
Страница 6: ......
Страница 7: ...CIRRUS photo CIRRUS photo 600 and CIRRUS photo 800 User manual...
Страница 8: ...000000 2121 813 GA GB 050416...
Страница 73: ......
Страница 76: ......
Страница 77: ...Remote maintenance tool Addendum to the documentation set...
Страница 78: ...000000 2121 813 AddGA01 GB 210915...
Страница 85: ......
Страница 88: ......
Страница 104: ...WINDOWS EMBEDDED STANDARD 7 GB 21 06 2012 16...
Страница 105: ......