Maintenance and care
000000-2121-813-GA-GB-050416
37
Maintenance and care
WARNING - GENERAL HAZARDS
Further maintenance procedures above and beyond those specified in this
section (maintenance, safety inspections and repairs) may only be carried
out by persons authorized by Carl Zeiss Meditec and solely according to the
service instructions issued by Carl Zeiss Meditec. For planning and
implementing these maintenance and care procedures please contact ZEISS
service or your local dealer.
Fault remedy
Fault
Possible cause
Remedy
No electric function
Power switch not switched on
Switch power switch to ON
Basic unit power cable not
connected
Connect cables to power supply
No image on monitor
Monitor power cable not connected
Connect cables to power module
(5, Fig.
Monitor displays message
"Cable not connected"
Monitor signal cable not connected
to basic unit
Connect monitor signal cable to
basic unit
No observation light
Rotary knob for brightness control
adjusted to minimum
Turn rotary knob to the right until
image is sufficiently visible
Observation brightness is not
continuously adjustable
Brightness control is defective
Contact service department
External fixation lamp is dark
Flashing diode defective
Replace flashing diode as described
in section
diode of the external fixation lamp
on page 38
Image is too dark, despite
high flash light level
Flash bulb defective
Contact service department
Содержание CIRRUS photo 600
Страница 1: ...CIRRUS photo CIRRUS photo 600 and CIRRUS photo 800 Documentation set...
Страница 4: ......
Страница 6: ......
Страница 7: ...CIRRUS photo CIRRUS photo 600 and CIRRUS photo 800 User manual...
Страница 8: ...000000 2121 813 GA GB 050416...
Страница 73: ......
Страница 76: ......
Страница 77: ...Remote maintenance tool Addendum to the documentation set...
Страница 78: ...000000 2121 813 AddGA01 GB 210915...
Страница 85: ......
Страница 88: ......
Страница 104: ...WINDOWS EMBEDDED STANDARD 7 GB 21 06 2012 16...
Страница 105: ......