OPERATION
Carl Zeiss
Axiovert 200 M
Axiovert 200
3-26
B 40-080 e 03/01
Deactivation of Set mode and saving of settings
The Set mode can be exited after conclusion of the settings. The settings / changes performed are
automatically stored when the mode is exited.
•
To exit the Set mode and return to the standard mode, briefly press the
Set LM
key on the
instrument rear.
The permanent storage of all the settings is confirmed by a beep.
Config.
& Focus Settings Stored
appears in the LCD display.
☞
If no operation is performed in the Set mode for 60 seconds, the microscope is automatically
reset to the standard mode. The settings made until then will not be stored.
If the instrument is switched off in the Set mode, the settings performed will not be stored
either.
3.2.2.2
Setting of focus limit (software limit)
In addition to the recognition of the hardware end positions, the software also permits the setting of the
focus limit, i.e. the upper limit of the travel range of the focusing drive.
This limit is an absolute focus position which applies to all nosepiece positions and which is only
corrected by the amount of parfocality alignment of the relevant objective.
In the factory alignment of the Axiovert 200 M, the focus limit lies above the upper end position switch
and is therefore not effective.
Activation of focus limits
•
Press
Focus Up
key for more than 2 seconds.
The current position of the focusing drive is stored as software limit (topmost Z-position).
FOC-Limit set
appears in the LCD display for 4 seconds.
Shifting of focus limit
•
Keep the
Work
key pressed and simultaneously move to the required position via the focusing
drive.
The new focus limit is stored after the key is released.
Содержание Axiovert 200
Страница 1: ...Operating Manual Axiovert 200 Axiovert 200 M Inverted Microscopes...
Страница 11: ...INTRODUCTION Axiovert 200 Overall view of the Axiovert 200 and Axiovert 200 M Carl Zeiss B 40 080 e 03 01 0 11...
Страница 16: ...INSTRUMENT DESCRIPTION Carl Zeiss Contents List of illustrations Axiovert 200 1 2 B 40 080 e 03 01...
Страница 30: ...INSTRUMENT DESCRIPTION Carl Zeiss Technical Data Axiovert 200 1 16 B 40 080 e 03 01...
Страница 34: ...START UP Carl Zeiss List of illustrations Axiovert 200 2 4 B 40 080 e 03 01...
Страница 108: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Carl Zeiss Contents List of illustrations Axiovert 200 4 2 B 40 080 e 03 01...
Страница 114: ...ANNEX Carl Zeiss Overview Axiovert 200 A 2 B 40 080 e 03 01...
Страница 120: ...ANNEX Carl Zeiss List of key words Axiovert 200 A 8 B 40 080 e 03 01...