OPERATION
Carl Zeiss
Illumination and contrasting techniques
Axiovert 200
3-42
B 40-080 e 03/01
•
Switch on the NHBO 103 fluorescence illuminator (3-24/
1
) via the power supply and allow it to heat
to operating temperature for approx. 15 minutes.
•
Select the FL reflector module in the reflector turret containing the required fluorescence filter
combination (depending on the excitation) and switch it on.
•
Press the
FL on/off
key (3-24/
6
) to open the fluorescence shutter. The opening of the shutter is
confirmed by a beep.
For the epi-fluorescence technique, one slider each with iris stop is used as aperture and luminous-field
diaphragm. However, since the aperture diaphragm is not visible in reflected light, the aperture
diaphragm slider must first be centered in the compartment of the luminous-field diaphragm and then
be inserted in the compartment for the aperture diaphragm.
The centering for the aperture diaphragm and the luminous-field diaphragm is performed in the same
way:
•
Insert an iris stop slider into the compartment (3-24/
3
) for the luminous-field diaphragm until stop.
•
Use the lever to close the diaphragm until it becomes visible in the field of view.
•
Then center the diaphragm using the two adjusting screws in the slider (with SW 3 mm ball-headed
screwdriver). Open the diaphragm until the entire field is uncovered.
•
Push the slider centered in this way into the compartment for the aperture diaphragm (3-24/
2
) until
stop.
•
Insert an additional iris stop slider into the compartment (3-24/
3
) for the luminous-field diaphragm.
•
Close the luminous-field diaphragm until it becomes visible in the field of view.
•
Use the two centering screws to center the luminous-field diaphragm in the edge of the field of view.
•
Either open the luminous-field diaphragm until it just disappears behind the edge of the field of view
or, if there is a risk of specimen bleaching, reduce it until it is visible in the field of view.
•
Finally, refocus on the specimen and optimize the collector position of the fluorescence illuminator.
For this purpose, use the knurled knob on the fluorescence illuminator to set the collector in such a
way that illumination of the field of view provided by the short-wave excitation reflector module is as
homogeneous as possible. Correction of the collector position is not required when modules with
long-wave excitation are used.
Содержание Axiovert 200
Страница 1: ...Operating Manual Axiovert 200 Axiovert 200 M Inverted Microscopes...
Страница 11: ...INTRODUCTION Axiovert 200 Overall view of the Axiovert 200 and Axiovert 200 M Carl Zeiss B 40 080 e 03 01 0 11...
Страница 16: ...INSTRUMENT DESCRIPTION Carl Zeiss Contents List of illustrations Axiovert 200 1 2 B 40 080 e 03 01...
Страница 30: ...INSTRUMENT DESCRIPTION Carl Zeiss Technical Data Axiovert 200 1 16 B 40 080 e 03 01...
Страница 34: ...START UP Carl Zeiss List of illustrations Axiovert 200 2 4 B 40 080 e 03 01...
Страница 108: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Carl Zeiss Contents List of illustrations Axiovert 200 4 2 B 40 080 e 03 01...
Страница 114: ...ANNEX Carl Zeiss Overview Axiovert 200 A 2 B 40 080 e 03 01...
Страница 120: ...ANNEX Carl Zeiss List of key words Axiovert 200 A 8 B 40 080 e 03 01...