OPERATION
Carl Zeiss
Axiovert 200
Axiovert 200
3-6
B 40-080 e 03/01
On / Off switch
(3-1/
1
)
−
position 0 = instrument switched off
−
position I = instrument switched on, switch lights up
Key HAL on/off
(3-1/
10
)
−
Brief pressing switches the halogen illuminator on or off alternately.
−
Long pressing (> 1 s) automatically sets the brightness to 3200K for color photography.
Toggle switch for brightness control of the halogen illuminator
(3-1/
11
)
−
Controls the brightness of the halogen illuminator in 2 steps.
step 1 – slow brightness change:
switch pressed half down until first pressure point:
switch to the left – brightness reduction
switch to the right – brightness increase
step 2 – fast brightness change:
switch pressed down until stop (second pressure point)
switch to the left – brightness reduction
switch to the right – brightness increase
−
Variable voltage range between 0 and 12 V. When the limit values are reached, a warning beep is
emitted.
−
At 10.5 V, the halogen lamp provides light with the color temperature of 3200 K for color
photography. To make this setting easier for you, the fast increase of the lamp voltage stops at this
point and a beep is emitted. To further increase the lamp voltage, keep the toggle switch pressed for
more than 1 sec
or
release it briefly and then press it again.
−
If the toggle switch is pressed down for more than 3 sec while the halogen illuminator is switched off,
the halogen illuminator will automatically switch on again.
Focusing drive
(3-1/
8
)
−
Coarse adjustment (big knob):
Focusing knob for coarse adjustment on both sides of the instrument: 1 coarse rotation = 5 mm
Overall lift: approx. 10 mm
−
Fine adjustment (small knob):
Focusing knob for fine adjustment available on both sides of the instrument: 1 fine rotation = 0.1 mm
Clamping lever for vertical stop of the focusing drive
(3-1/
33
)
−
Setting of vertical stop for stage focusing.
−
Turn clamping lever for stop upwards towards the stopping pin. Use the focusing drive to move the
stage to the uppermost admissible position where collisions with the stage or the microscope slide are
no longer possible. Then press the clamping lever downwards to clamp it in the stop position.
Содержание Axiovert 200
Страница 1: ...Operating Manual Axiovert 200 Axiovert 200 M Inverted Microscopes...
Страница 11: ...INTRODUCTION Axiovert 200 Overall view of the Axiovert 200 and Axiovert 200 M Carl Zeiss B 40 080 e 03 01 0 11...
Страница 16: ...INSTRUMENT DESCRIPTION Carl Zeiss Contents List of illustrations Axiovert 200 1 2 B 40 080 e 03 01...
Страница 30: ...INSTRUMENT DESCRIPTION Carl Zeiss Technical Data Axiovert 200 1 16 B 40 080 e 03 01...
Страница 34: ...START UP Carl Zeiss List of illustrations Axiovert 200 2 4 B 40 080 e 03 01...
Страница 108: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Carl Zeiss Contents List of illustrations Axiovert 200 4 2 B 40 080 e 03 01...
Страница 114: ...ANNEX Carl Zeiss Overview Axiovert 200 A 2 B 40 080 e 03 01...
Страница 120: ...ANNEX Carl Zeiss List of key words Axiovert 200 A 8 B 40 080 e 03 01...