OPERATION
Axiovert 200
Axiovert 200 M
Carl Zeiss
B 40-080 e 03/01
3-19
LCD display
(3-13/
17
)
−
Two-line display with 16 characters each
attached to the carrier for transmitted-light
illumination / condenser.
−
Display of objective position (3-6/
1
),
magnification (3-6/
2
) and contrasting technique
(3-6/
3
) in the upper line.
−
FL
display (3-6/
4
) on the right of the top line if
the fluorescence shutter is opened.
−
Display of lamp brightness in a line scale and
display of the relevant voltage (3-6/
5
) in the
lower line. One interval corresponds to 0.4 V.
−
The LCD background illumination can be
switched on or off as required. Activation /
deactivation is performed by pressing the
FL
key for approx. 1 second.
−
Brightness adjustment of the LCD display is by
pressing/holding the
FL
key + the
toggle
switch
.
−
The LCD display is also used for the
configuration of the microscope (Set mode).
Instead of the line scale for the lamp brightness, the following can be displayed in the lower line (3-6/
6
)
of the LCD display:
−
HAL off
when the halogen illuminator is switched off.
−
3200K
when the lamp voltage has been set to the color temperature (10.5 V).
−
Description of the swung-in filter set on the reflector turret (e.g.
DAPI/FITC
), provided that the
fluorescence shutter has been opened. If no filter set is contained in the reflector turret, the reflector
turret position (e.g.
Refl. 2
) is displayed instead.
The currently set position appears in the LCD display for 4 seconds during performance of the following
key functions: Both lines of the display are used.
−
1x Optovar
: display of the Optovar magnification factor (
1.6x
) and the overall magnification (
Total
Mag.
) in the intermediate image
−
LEFT/RIGHT
and
Slider VIS/Frontport/Baseport
: display of the actual intensities at Sideport,
Frontport, Baseport or VIS, depending on the selected positions of both control elements, e.g. SP:
Right 20%
(upper line of the LCD display) and
Binocular 80%
(lower line).
−
FOCUS - Load
:
Load Position
appears in the lower line (display blinks every second).
−
ZERO: display of the current focus position in mm (e.g.
Zdis: +0.0µm
)
−
If a motorized condenser is connected, the currently set condenser turret position appears in the
lower line (e.g.
Cond. DICIII
). If the aperture is changed by pressing the key on the condenser, the
value of the stop opening is displayed (e.g.
Cond. NA: 0.55
). When a phase stop position is set, the
aperture diaphragm is always entirely open.
−
Display function during the instrument configuration (SET mode).
Fig. 3-14
LCD-Display
Содержание Axiovert 200
Страница 1: ...Operating Manual Axiovert 200 Axiovert 200 M Inverted Microscopes...
Страница 11: ...INTRODUCTION Axiovert 200 Overall view of the Axiovert 200 and Axiovert 200 M Carl Zeiss B 40 080 e 03 01 0 11...
Страница 16: ...INSTRUMENT DESCRIPTION Carl Zeiss Contents List of illustrations Axiovert 200 1 2 B 40 080 e 03 01...
Страница 30: ...INSTRUMENT DESCRIPTION Carl Zeiss Technical Data Axiovert 200 1 16 B 40 080 e 03 01...
Страница 34: ...START UP Carl Zeiss List of illustrations Axiovert 200 2 4 B 40 080 e 03 01...
Страница 108: ...CARE MAINTENANCE TROUBLESHOOTING AND SERVICE Carl Zeiss Contents List of illustrations Axiovert 200 4 2 B 40 080 e 03 01...
Страница 114: ...ANNEX Carl Zeiss Overview Axiovert 200 A 2 B 40 080 e 03 01...
Страница 120: ...ANNEX Carl Zeiss List of key words Axiovert 200 A 8 B 40 080 e 03 01...