Axioskop 2 FS
plus
OPERATION
Axioskop 2 FS
MOT
Operation and function controls
Carl Zeiss
B 40-076 e 02/01
3-15
Display
The set step width or the positive or negative
position deviation with reference to the zero point
is shown in the display field of the keypad. When
setting functions are performed, the display is used
to show the changeable setting values.
Handwheel for fine adjustment
The handwheel permits the very sensitive
adjustment of the focusing drive (minimum:
0.1 µm) for very precise focus settings at high
objective magnifications in particular. Turning the
handwheel moves the objective slider upwards or
downwards via the focusing drive.
Upwards / downwards direction keys
The
upwards / downwards
direction keys
control the upward and downward movement of
the focusing drive. When one of the direction keys
is pressed, the setting of the focusing drive is
changed at the preset speed until the key is
released again.
The speed of the focusing drive can be roughly set to the
slow
and
fast
speed ranges via the
Slow
and
Speed
keys. The relevant speed range LED integrated in the left upper corner of the key lights up.
It is also possible to set the speed of the focusing drive for the
Slow
and
Speed
ranges independently of
each other in 16 steps. This allows the single-hand control to be easily matched to individual
requirements.
The steps for the Slow and Speed ranges are set as follows:
•
Press the key of the speed range to be changed
(e.g. Slow; the LED of the Slow key lights up).
•
Press the
Set Speed Slow
key. The current
speed step is shown in the display of the
keypad.
Fig. 3-12
Display
Fig. 3-13
Handwheel for fine adjustment
Fig. 3-14
Display-appearing when speed is
set in the Slow range
Содержание Axioskop 2 FS mot
Страница 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Страница 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Страница 36: ...START UP Axioskop 2 FS plus Carl Zeiss List of illustrations Axioskop 2 FS MOT 2 4 B 40 076 e 02 01...
Страница 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Страница 112: ...CARE MAINTENANCE TROUBLESHOOTING Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 4 2 B 40 076 e 02 01...
Страница 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...