START-UP
Axioskop 2 FS
plus
Carl Zeiss
Initial start-up
Axioskop 2 FS
MOT
2-10
B 40-076 e 02/01
2.1.7
Insertion of DIC slider
If the microscope equipment was ordered for applications in differential interference contrast, the DIC
sliders are already inserted in the objective sliders and centered.
For retrofitting, or to change the DIC slider, proceed as follows:
•
Remove the stage carrier (if not yet done - see
2.1.4) and safely place it on the table.
•
Bring front objective mount in the work
position (see page 3-8).
•
Push suitable DIC slider (2-8/
1
, see labels) in the
provided opening (2-8/
2
) of the objective slider
above the objective.
•
Make sure that the DIC slider precisely locks
into position.
•
Bring rear objective mount in the work position
(see page 3-8) and proceed in the same way as
described above.
☞
After being inserted in the objective slider, the DIC sliders must be moved into center position.
However, this requires the microscope to be put in operation completely and the KÖHLER
illumination to be set (see chapter 3, section 3.3.1). Setting of the DIC sliders to the center
position is described in section 3.3.4 (3) of chapter 3.
•
When the objectives, the DIC slider and the condenser have been inserted, the stage carrier must be
attached to the microscope again (see section 2.1.8).
Fig. 2-8
Insertion of DIC slider
Содержание Axioskop 2 FS mot
Страница 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Страница 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Страница 36: ...START UP Axioskop 2 FS plus Carl Zeiss List of illustrations Axioskop 2 FS MOT 2 4 B 40 076 e 02 01...
Страница 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Страница 112: ...CARE MAINTENANCE TROUBLESHOOTING Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 4 2 B 40 076 e 02 01...
Страница 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...