B 40-076 e 02/01
2-33
☞
To ensure long life of the stage, the scuff from the specimen holders should be removed at
regular intervals. Make sure that the scuff does not get into the guiding components for the x-
movement.
(6)
Changing the universal mounting
frame
•
Loosen two countersunk screws (2-45/
1
) using
SW 2 Allen key (2-45/
3
).
•
Remove universal mounting frame (2-45/
2
)
from the mechanical stage (2-45/
4
) in upward
direction.
•
Attach required mounting frame or specimen
holder (2-45/
5
) to the mechanical stage and
tighten the two countersunk screws.
2.2.10
Changing the binocular tube
•
Use SW 3 ball-headed screwdriver to loosen
clamping screw (2-46/
3
) and remove available
tube in an upward direction.
•
Place dust cap (2-46/
2
) for tube lens protection
above the dovetail of the binocular tube.
•
Remove dust cap from the required tube.
•
Insert dovetail of tube (2-46/
1
) in the stand
opening (2-46/
4
) and align the tube.
•
Tighten clamping screw (2-46/
3
).
Fig. 2-45
Attaching the universal mounting
frame
FS
Fig. 2-46
Changing the binocular tube
Содержание Axioskop 2 FS mot
Страница 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Страница 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Страница 36: ...START UP Axioskop 2 FS plus Carl Zeiss List of illustrations Axioskop 2 FS MOT 2 4 B 40 076 e 02 01...
Страница 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Страница 112: ...CARE MAINTENANCE TROUBLESHOOTING Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 4 2 B 40 076 e 02 01...
Страница 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...