6 On Screen Display (OSD) menu | 6.3 Settings menu
ZEISS
Applying Bright Pixel Correction
Prerequisite
ü
The C-mount port is closed so that no light will reach the image sensor during the procedure.
You can close the C-mount port by either closing the light path of the microscope stand or
detaching the camera from the stand and protect sensor from light by screwing the protective
cap onto the camera's c-mount thread.
1. Tap the
Correct
button.
à
Do not operate the camera during the procedure.
à
If light reaches the sensor, the procedure fails.
Either cover the camera properly or reset the Bright Pixel configuration before restarting
the procedure.
The Bright Pixel Correction procedure is applied.
The camera reboots after a successful procedure.
User Guide ZEISS Axiocam 208 color / 202 mono | en-US | Rev.1 | 426570-7011-001
41