Image Acquisition
Axiocam 105 color Instal Reference Guide
23
Automatic Exposure
The
Auto Snap
feature is not available for the Axiocam 105 color.
If the
Auto Live
check box is activated, the software continuously sets the
exposure during the live image. In this way it is possible to generate
images with the correct exposure without having to manually enter an
exposure time. The exposure time currently calculated in the live image is
used for each individual image.
The exposure measurement button
can be pressed at any
time. The Auto Live function will then be deactivated. This is useful if you
want to update the exposure time for a current live image.
Note
:
In this mode it is not possible to edit the exposure times in the
multichannel mode for the individual channels.
White Balance
In order to achieve a color display which is neutral and close to reality it is
necessary to balance the individual color channels against one another.
This requires a scene without a sample to be set in the microscope. The
scene should be white, neutral and illuminated as evenly as possible. It
must not be overexposed.
: Using this button you can specify a reference point for
the white balance yourself. To do this, use the pointer to click on a
white area in the live image.
: Starts the measurement. The weighting factors
calculated for the color channels are then displayed.
: Using this button you can switch back to the basic white
balance setting. This setting is optimized for a halogen light
temperature of 3200 K.
Содержание Axiocam 105 color
Страница 1: ...Axiocam 105 color Installation Reference 10 2013 October 2013 ...
Страница 2: ......
Страница 6: ......
Страница 18: ...Installation 10 Printed 10 2013 ...