CLEANING, maintenance and service
Carl Zeiss
Maintenance
Axio Zoom.V16
58
435080-9030-701
05/2012
5.2
Maintenance
5.2.1
Safety inspections
When the motorized focusing drive is lowered there is the
risk of pinching fingers
in the
working area.
Perform regular checks of the travel range of the motorized focusing drive to ensure that the
limit switches are working properly in accordance with Section 4.6, page 44.
5.2.2
Relocation of the device
Do not transport the installed, upright standing system over long distances. Avoid severe shocks
to the focusing drive, as this could damage the rack of the column.
For transporting over long distances dismantle the instrument and ship it in the original
packaging.
To relocate the instrument over short distances, e.g. within a hospital, proceed as follows:
•
Move the focusing drive to the lower stop.
•
Switch the instrument off and pull all cable connections.
•
Dismantle the phototube, spacer tube, objective(s) and microscope body. For this purpose see Section
3.3, page 20 (reverse order for dismantling).
•
If it has been used, dismantle the transmitted light unit S or transillumination top 450 mot. For this
purpose see Section 3.9, page 31 (reverse order for dismantling).
Please note that the center of gravity is located above the stand base. For this reason the
instrument should not be carried over long distances. Do not tilt or overturn the instrument.
Use the recessed grips on the edge of the stand base to lift and set down the instrument.
•
Lift the instrument by the recessed grips on the stand base, not by taking hold of the focusing drive.
•
Re-assemble components, connect the instrument as specified and switch it on.