Changing insert plate 450
CAUTION!
Risk of pinching fingers!
When inserting or removing the ferromagnetic insert plate 450 or the heatable ferromagnetic
insert plate 450, use the supplied flat plastic material to avoid pinching fingers.
The controller SC 300 must be switched off before installing or removing the heatable
ferromagnetic insert plate 450.
Let the heatable ferromagnetic insert plate 450 cool down to room temperature before
removing it.
CAUTION!
Hot surface!
ATTENTION!
Danger of breakage!
The glass insert plate 120 mm is only 1.1 mm thick and thus fragile. It must be handled with
care! Do not apply pressure to the glass. Hold it only by the metal frame!
The following standard options are available as an alternative to the insert plate 450:
−
Ferromagnetic insert plate 450
−
Heatable insert plate 450 as part of heating unit 450
•
Remove glass or B/W plate (Fig. 20/
8
) and insert 52/84 (Fig. 20/
7
).
•
Remove the base insert plate (Fig. 20/
3
). To do this, loosen the short socket head screws (Fig. 20/
2
).
•
Install the ferromagnetic insert plate 450 (Fig. 20/
10
) or heatable insert plate 450 and secure with
three socket head screws.
If the heatable ferromagnetic insert plate 450 is used, pass the two heating cables side-by-side
through the cut-out in the insert plate and secure with three clamps. Then pass the two heating
cables to the outside through the lateral cavity (Fig. 20/
11
) in the transmitted light unit S and
connect to controller SC 300.
For installation and operation of control unit SC 300 please refer to the separately delivered
operating manual.