INSTALLATION INSTRUCTIONS AND FIRST-TIME SET-UP
Axio Observer
HAL 100 illuminator
ZEISS
12/2016
431004-7244-001
67
4.22.3
Replacing the halogen bulb 12 V
100 W
Hot surface!
The lamp housing does not need to be
removed from the stand in order to
replace the halogen bulb. The bulb
replacement tool (Fig. 53/
7
) is
not
to
be stored in the lamp housing when
the illuminator is in use.
The spare bulb (Fig. 53/
8
) can remain
in place in the lamp housing.
•
Switch off the microscope and remove the plug
(Fig. 50/
6
) from the 12 V/100 W socket
3
for reflected light or Fig. 50/
5
for
transmitted light). Allow to cool for about 15
min.
•
Press the release button (Fig. 53/
3
) on the
HAL 100 illuminator (Fig. 53/
1
) downwards and
remove the bulb holder (Fig. 53/
2
) completely.
Place it on a flat surface.
•
Press down the two spring levers (Fig. 53/
5
) and
remove the old halogen bulb (Fig. 53/
6
) by
pulling it upwards.
•
Press down the two spring levers and insert the
new bulb into the bulb socket (Fig. 53/
4
).
Release the spring levers. Always hold the
halogen bulb by means of the bulb replacement
tool (Fig. 53/
7
), as traces of grease on the bulb
may shorten bulb life.
•
Briefly depress the spring levers once more to
center the bulb.
•
Reinsert the bulb holder until you feel it click
into place.
Fig. 53
Replacing the halogen bulb