2.1 Preventive
Maintenance
Scheduled preventative maintenance visits are not required for the ATLAS 9000. Any routine
operator maintenance (if needed) is listed in the ATLAS 9000 User Manual. For preventative
maintenance requirements on any of the optional peripheral equipment, please refer to the
related manufacturer’s owner/user manual.
Basic troubleshooting tasks the operator can perform are listed below:
•
Check that all external cable and cord connections are secure.
•
Check that the instrument is being powered from a properly grounded AC outlet.
•
Observe the ambient conditions in which the instrument is operating.
Make sure
that the instrument is not used in direct sunlight !
•
Are the cooling vents on the instrument unobstructed?
•
Are there any ambient conditions present that may case static generation (carpeted
floors, excessively dry weather, or other causes)?
•
Ensure that all instrument precautions (Section 1.4) are being observed.
2.2 System
Checkout
Procedure
The system checkout procedures are used to verify that the instrument is operating properly
within specifications. System checkout must be performed at the completion of every
installation and service call on the instrument. If an error is detected during system checkout,
troubleshoot and repair the instrument; then start the system checkout again from the
beginning. The entire system checkout should always be performed, from start to finish,
without any problems occurring.
The checklist includes several checks that are essential following any service of the instrument.
The following special tools/equipment are required to successfully complete the System
Checkout Checklist:
•
ATLAS Test Eye Tool
Service Representatives in the United States must perform the System Checkout procedure
and complete the Service Checklist form on each service call. International Service
Representatives should forward their completed checklists to the local Zeiss affiliate or
distributor.
Содержание ATLAS 9000
Страница 4: ......
Страница 10: ...General Information ATLAS Model 9000 Confidential and Proprietary Service Manual 1 2 63041A0208 Notes...
Страница 26: ...Parts Removal Replacement ATLAS Model 9000 Confidential and Proprietary Service Manual 3 2 63041A0208 Notes...
Страница 38: ...Adjustment and Calibration ATLAS Model 9000 Confidential and Proprietary Service Manual 4 2 63041A0208 Notes...
Страница 44: ...Troubleshooting ATLAS Model 9000 Confidential and Proprietary Service Manual 5 2 63041A0208 Notes...
Страница 49: ...Diagrams ATLAS Model 9000 Confidential and Proprietary Service Manual 6 2 63041A0208 Notes...
Страница 51: ...FIGURE 6 2 ATLAS Model 9000 System Wiring Diagram...
Страница 53: ...Parts ATLAS Model 9000 Confidential and Proprietary Service Manual 7 2 63041A0208 Notes...
Страница 60: ...ATLAS Model 9000 Parts Confidential and Proprietary 63041A0208 7 9 Service Manual 7 7 1 TFT Assembly...
Страница 62: ...ATLAS Model 9000 Parts Confidential and Proprietary 63041A0208 7 11 Service Manual 7 7 2 Tower Assembly...
Страница 64: ...ATLAS Model 9000 Parts Confidential and Proprietary 63041A0208 7 13 Service Manual 7 7 3 Y Drive Assembly...
Страница 66: ...ATLAS Model 9000 Parts Confidential and Proprietary 63041A0208 7 15 Service Manual 7 7 4 X Z Stage Assembly...
Страница 72: ...Appendices ATLAS Model 9000 Confidential and Proprietary Service Manual Appendices 2 63041A0208 Notes...
Страница 76: ...Appendix A ATLAS Model 9000 Confidential and Proprietary Service Manual A 4 63041A0208...
Страница 112: ...Appendix D ATLAS Model 9000 Confidential and Proprietary Service Manual D 2 63041A0208...
Страница 114: ...Appendix E ATLAS Model 9000 Confidential and Proprietary Service Manual E 2 63041A0208...
Страница 116: ...Appendix F ATLAS Model 9000 Confidential and Proprietary Service Manual F 2 63041A0208...