Instruction for use
7 Maintenance
AT 030
7.2 Adjusting the swivel-mounted tonometer holder with square mounting
319300-9120-000-GA-GB-240320
37 / 60
If the same eyepiece is always to be used for tonometry, the
adjustment is now finished.
If, however, both eyepieces are to be used alternately, the position
of the AT 030 on the base plate must now be changed:
u
Release the knurled screw and remove the tonometer.
u
Set the tonometer with the centering pin into the other
alignment hole of the base plate.
u
Retighten the knurled screw.
u
Check, whether the measuring prism is also correctly aligned in
this eyepiece. If not, repeat the first four steps.
Setting/checking the correct focus position (image focus C)
u
Set the scale on the tonometer to zero.
u
To assess the focus sharpness apply a visible structure to one
side of the front surface of the measuring prism (e.g. finger-
print, dust, graph paper).
Содержание AT 030
Страница 1: ...AT 030 Applanation tonometer Documentation set...
Страница 4: ......
Страница 6: ......
Страница 7: ...AT 030 Applanation tonometer Instruction for use...
Страница 8: ......
Страница 16: ...Empty page for your notes...
Страница 20: ...Empty page for your notes...
Страница 28: ...Empty page for your notes...
Страница 36: ...Empty page for your notes...
Страница 38: ...Empty page for your notes...
Страница 52: ...Empty page for your notes...
Страница 56: ...Empty page for your notes...
Страница 60: ...Empty page for your notes...
Страница 62: ...Empty page for your notes...
Страница 64: ...Empty page for your notes...
Страница 68: ......
Страница 69: ...Instructions for use and disinfection of tonometer measuring prisms Addendum...
Страница 70: ......
Страница 72: ...Empty page for your notes...
Страница 85: ...Empty page for your notes...
Страница 87: ...Empty page for your notes...