7 Maintenance
Instruction for use
AT 030
7.2 Adjusting the swivel-mounted tonometer holder with square mounting
interface (version a)
34 / 60
319300-9120-000-GA-GB-240320
Lateral and vertical deviations of 1 mm do not impact the pressure
measurement precision.
In the case of greater deviations, first check the following:
Action
1. Is the tonometer positioned in front of the "correct" eyepiece?
2. Is the tonometer holder properly mounted?
3. Has the tonometer holder properly engaged in the operating
position?
If these checks do not yield satisfying results, align tonometer
holders which are mounted behind the stereomicroscope, as
described in section Adjusting the swivel-mounted tonometer
holder with square mounting interface (version a) [
Adjusting the swivel-mounted tonometer holder with round
mounting interface (version b) [
39]. It is not necessary to remove
the tonometer holder with the AT 030 mounted from the slit lamp.
NOTE! If, in case of the swivel-mounted tonometer holder
mounted on the stereomicroscope or Fundus VarioView and
in case of fixed tonometer holders, these checks do not yield
satisfying results, contact ZEISS Service.
Due to manufacturing tolerances (or wear) on the slit lamp it is
possible that, although the tonometer holder has been adjusted
accurately on one side, after shifting to the other side, the
adjustment is unsatisfactory. In this case, repeat adjustments alter-
nately on both sides until it is optimally adjusted in both positions.
NOTE! The adjustment steps can be carried out indepen-
dently of each other in any order. If adjustment according to
the above instructions is not successful, please contact
ZEISS Service.
7.2 Adjusting the swivel-mounted tonometer holder
with square mounting interface (version a)
Preparation
Action
u
Mount the tonometer holder and AT 030 on the slit lamp.
u
Put the AT 030 into the position for the right eyepiece.
u
Set the eyepieces of the slit lamp to 0 D or your own correction
value.
u
Move the AT 030 to the measuring position.
u
Illuminate the measuring prism with the slit lamp projector
from the left to approx. 90°.
Содержание AT 030
Страница 1: ...AT 030 Applanation tonometer Documentation set...
Страница 4: ......
Страница 6: ......
Страница 7: ...AT 030 Applanation tonometer Instruction for use...
Страница 8: ......
Страница 16: ...Empty page for your notes...
Страница 20: ...Empty page for your notes...
Страница 28: ...Empty page for your notes...
Страница 36: ...Empty page for your notes...
Страница 38: ...Empty page for your notes...
Страница 52: ...Empty page for your notes...
Страница 56: ...Empty page for your notes...
Страница 60: ...Empty page for your notes...
Страница 62: ...Empty page for your notes...
Страница 64: ...Empty page for your notes...
Страница 68: ......
Страница 69: ...Instructions for use and disinfection of tonometer measuring prisms Addendum...
Страница 70: ......
Страница 72: ...Empty page for your notes...
Страница 85: ...Empty page for your notes...
Страница 87: ...Empty page for your notes...