Process Variable
Type
Values
Description
UtilErrReport stringin
command
string
Returns last command to cause an error.
Note:
This functionality is only valid for the ZT44xx
UtilErr
stringin error string
Error string explaining current error code
UtilTemp
ai
0.0 to 100.0 Internal temperature in degrees C
UtilMem longin
Varies
by
model, see
instrument
manual
Waveform capture memory in samples per
channel, varies by installed memory option
UtilMemClear
bo
0 or 1
Clears all non-volatile system memory. A transition
from 0 to 1 causes the clear, after which the value
returns to 0.
UtilSelfCal
longin
0,1 or 2
Initiates and returns the result of the unit self-
calibration process. The internal calibration
determines the zero DC offset, the DC offset
adjust scale factor, and the ADC balance for all
input range settings for all input channels. The
internal calibration process can take several
minutes to complete. The instrument is reset upon
completion of the calibration process.
Note:
The input channels must be disconnected
or be driven with 0.0 VDC before starting the
calibration.
The timeout value should be set to infinite before
starting the calibration, and reset to the default
value when completed.
Note:
Do not interrupt the instrument during
calibration or the calibration tables could be
corrupted.
0 Pass
1 Fail (Did not converge)
2 Corrupt
setUtilSelfID
getUtilSelfID
bo
bi
0 or 1
Instrument HST LED state:
0 = LAN status
1 = blink
Allows instrument identification.